强激光与粒子束, 2001, 13 (5): 565, 网络出版: 2006-05-15   

透射光栅谱仪的整体标定

Integral calibration for transmission grating spectrometer
作者单位
1 中国工程物理研究院激光聚变研究中心,四川,绵阳,621900
2 中国科学院高能物理研究所同步辐射实验室,北京,100039
摘要
提出了透射光栅谱仪整体标定的建议,并在北京同步辐射源上首次对ICF实验中所用的透射光栅谱仪进行了整体标定,获得了与分别标定基本一致的结果。与分别标定相比,整体标定具有实验步骤少、实验数据处理简单、标定结果使用更加直接等优点。
Abstract
The concept of integral calibration for the transmission grating spectrometer is suggested.Compared with other calibration method,the integral calibration method is more convenient and valid,and its data process is simpler.Also,the calibration results can be more directly used.The calibrated experiments were first carried out on Beijing Synchrotron Radiation Facility (BSRF).The experimental results agree well with the calibrated results for transmission grating spectrometer and X ray CCD separately.

曹磊峰, 江少恩, 杨家敏, 成金秀, 郑志坚, 温天舒, 崔明启, 朱佩平, 赵屹东, 黎刚. 透射光栅谱仪的整体标定[J]. 强激光与粒子束, 2001, 13(5): 565. 曹磊峰, 江少恩, 杨家敏, 成金秀, 郑志坚, 温天舒, 崔明启, 朱佩平, 赵屹东, 黎刚. Integral calibration for transmission grating spectrometer[J]. High Power Laser and Particle Beams, 2001, 13(5): 565.