Non-invasive Measurement of Simulated Burn Depth Using Intensity Fluctuations of Diffusion Light

作者单位
1 Institute of Laser Life Science, South China Normal University, Guangzhou 510631, China
2 Department of Applied Physics, Tianjin University, Tianjin 300072, China
摘要
Abstract
The determination of burn depth is an important problem for surgeons to remedy severe skin burns. The burn can be simplified as a two-layer media model: a static layer and a dynamic layer. In this paper, the relative fluctuation of backscattering light intensity was used to determine the simulated burn depth. The experimental results show that the relative fluctuation of light intensity increases with the simulated burn depth and the depth less than 0.1 mm can be resolved. This new method may be useful for burn diagnosis.

ZHANG Xiqin, XING Da, LIU Ying, MA Shining. Non-invasive Measurement of Simulated Burn Depth Using Intensity Fluctuations of Diffusion Light[J]. Chinese Journal of Lasers B, 2001, 10(4): 317. ZHANG Xiqin, XING Da, LIU Ying, MA Shining. [J]. 中国激光(英文版), 2001, 10(4): 317.