首页 > 论文 > 激光与光电子学进展 > 47卷 > 3期(pp:31101--1)

无机械扫描谱域光学相干层析术研究

Research on Non-Mechanical Scan Spectral-Domain OCT

  • 摘要
  • 论文信息
  • 参考文献
  • 被引情况
  • PDF全文
分享:

摘要

为提高光学相干层析术(OCT)系统成像速度,借助成像光谱仪,以钨灯为非相干光源,采用柱面镜聚焦,搭建了迈克耳孙干涉仪结构的无扫描装置谱域OCT系统。介绍了系统结构和工作原理,对待测样品进行了层析成像实验,分析了系统的成像特性,利用中值滤波的方法消除背景噪声,获得了样品的二维层析图像,与点扫描光纤型OCT的成像速度进行了比较。研究表明,无扫描装置谱域OCT系统在8 mm宽度范围成像时,其数据采集速度比传统点扫描方式约快2个数量级。

Abstract

In order to overcome the disadvantage of slow mechanical scan of optical coherence tomography (OCT), a non-mechanical scan frequency-domain OCT system with imaging spectrometer and Michelson interferometer was set up. The tungsten lamp was used as a non-coherent broadband light source. The focused line was obtained by using cylinder lens. The system structure and work principle were introduced respectively. The experiment to get optical tomography imaging of a sample was carried out. The imaging characteristic of the system was analyzed. The background noise is reduced by using the method of median filtering average. The result shows that the data collecting speed of the system is one hundred times as fast as that of time-domain tomography imaging in scope of 8mm, although the image quality is not as good as that of time-domain OCT due to the light source.

Newport宣传-MKS新实验室计划
补充资料

中图分类号:TN247

DOI:10.3788/lop47.031101

所属栏目:成像系统

基金项目:陕西省教育厅专项科研计划项目(04JK238),西北大学研究生创新基金(07YJC08)资助课题。

收稿日期:2009-08-02

修改稿日期:2009-10-19

网络出版日期:0001-01-01

作者单位    点击查看

忽满利:西北大学 物理系, 陕西 西安 710069
孙创伟:西北大学 物理系, 陕西 西安 710069
高爱华:西北大学 物理系, 陕西 西安 710069
张群喜:陕西省历史博物馆, 陕西 西安 710061

联系人作者:忽满利(huml@nwu.edu.cn)

备注:忽满利|教授|主要从事信息光学与光电检测等方面的研究|(1959-),男,博士。

【1】D. Huang , E. A. Swanson , C. P. Lin et al .. Optical coherence tomography[J ]. Science, 1991, 254 (5035): 1178~1181

【2】Jing Zhijun, Jiang Jingying, Zhang Fan et al.. Design of CCD grating spectrometer with refrigeration and application in Fourier-domain OCT[J]. Science Technology and Engineering, 2006, 24(6): 3820~3822
经志军, 蒋景英, 张帆 等. 带制冷的CCD光栅光谱仪设计及其在频域OCT中的应用[J]. 科学技术与工程, 2006, 24(6): 3820~3822

【3】Li Gang, Ren Zhao, Lin Ling et al.. Study on the feasibility and optical imaging properties of high speed line-focused OCT[J]. Chin. J. Biomed. Eng., 2007, 26(1): 89~93
李刚, 任钊, 林凌 等. 高速线扫描OCT的可行性与光学成像特性的研究[J]. 中国生物医学工程学报, 2007, 26(1): 89~93

【4】B. Grajciar, M. Pircher, A. F. Fercher et al.. Parallel Fourier domain optical coherence tomography for in vivo measurement of the human eye[J]. Opt. Express, 2005, 13(4): 1131~1137

【5】Haida Liang, Borislava Peric, Michael Hughes. Optical coherence tomography for art conservation & archaeology[C]. SPIE, 2007, 6618: 661805

【6】Lin Ling, Li Gang, Zhang Taishi et al.. Calibration of depth-coordinate of spectral domain OCT image[J]. Chin. J. Sci. Instrum., 2008, 29(3): 466~471
林凌,李刚, 张泰石 等. 频域OCT图像的深度坐标标定[J]. 仪器仪表学报, 2008, 29(3): 466~471

引用该论文

Hu Manli,Sun Chuangwei,Gao Aihua,Zhang Qunxi. Research on Non-Mechanical Scan Spectral-Domain OCT[J]. Laser & Optoelectronics Progress, 2010, 47(3): 031101

忽满利,孙创伟,高爱华,张群喜. 无机械扫描谱域光学相干层析术研究[J]. 激光与光电子学进展, 2010, 47(3): 031101

您的浏览器不支持PDF插件,请使用最新的(Chrome/Fire Fox等)浏览器.或者您还可以点击此处下载该论文PDF