光子学报, 2009, 38 (11): 2853, 网络出版: 2010-05-10   

空间调制干涉光谱成像仪光谱定标技术研究

Wavelength Calibration of Spatially Modulated Imaging Fourier Transform Spectrometer
作者单位
1 中国科学院西安光学精密机械研究所, 西安 710119
2 中国科学院研究生院, 北京 100049
3 中国科学院光电研究院, 北京100190
摘要
为了对干涉光谱成像仪进行光谱定标, 针对空间调制该仪器的原理, 得出了其光程差和光谱分辨率的计算方法, 并分析了仪器的线型函数.在实验中使用多种单色光源(波长范围:450~950 nm)和扩束准直镜(焦距120 mm、口径50 mm)进行光谱定标测试.结果表明:影响实验室光谱定标不确定度的主要因素为d/f的测量误差和标准光谱的误差.
Abstract
Spatially Modulated Imaging Fourier Transform Spectrometer (SMIFTS) is an instrument depended on interference, which can obtain interferential information and achieve spectral information by fourier transform.Wavelength calibration is prerequisite for SMIFTS to obtain the spectrum information exactly.The way of extreme search that uses multi-lasers(wavelength range:450~950 nm) and collimating lens is carried out for the wavelength calibration of SMIFTS.Principle,method and result of the experiment are given then.The results indicate that the primary factors influencing lab wavelength calibration′s precision are two reasons:measure precision of d/f and the precision of standard spectrum.高静|gaojing@opt.ac.cn

高静, 计忠瑛, 崔燕, 石大莲, 周锦松, 相里斌, 王忠厚. 空间调制干涉光谱成像仪光谱定标技术研究[J]. 光子学报, 2009, 38(11): 2853. GAO Jing, JI Zhong-ying, CUI Yan, SHI Da-lian, ZHOU Jin-song, XIANG-LI Bin, WANG Zhong-hou. Wavelength Calibration of Spatially Modulated Imaging Fourier Transform Spectrometer[J]. ACTA PHOTONICA SINICA, 2009, 38(11): 2853.

本文已被 2 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。