Chinese Optics Letters, 2010, 8 (11): 1053, Published Online: Dec. 3, 2010
Diffraction grating single-shot correlation system for measurement of picosecond laser pulses Download: 512次
单次相关 衍射光栅 皮秒激光脉冲 050.1950 Diffraction gratings 120.4640 Optical instruments 190.2620 Harmonic generation and mixing
Abstract
We study and experimentally demonstrate a sensitive single-shot correlation system in which only a diffraction grating is used to produce a transverse time delay (TTD) in the reference pulse. The mechanism of the TTD introduced by the grating and the formation of the relative time delay (RTD) in the noncollinear correlation system are analyzed in detail. By using our system, we successfully measured the temporal duration of picosecond laser pulses, and a time resolution of ~0.047 ps is obtained at 1047 nm. The impact of the grating dispersion and the second harmonic beam walk-off effect on the measurement are considered.
Fuling Zhang, Xiaoping Ouyang, Meizhi Sun, Qunyu Bi, Xinglong Xie, Zunqi Lin. Diffraction grating single-shot correlation system for measurement of picosecond laser pulses[J]. Chinese Optics Letters, 2010, 8(11): 1053.