光学技术, 2010, 36 (2): 200, 网络出版: 2011-02-21   

基于Lissajous图技术的波长移相标定方法

Wavelength-tuned phase-shifting calibration based on the Lissajous figures technique
作者单位
南京理工大学 电光学院, 江苏 南京 210094
摘要
波长移相干涉仪可应用于大口径光学元件的检测。研究了基于Lissajous图技术的波长移相标定方法, 实现了干涉图间移相量的计算, 解决了在不同干涉腔长下进行测试时移相量需控制成为一个特定值的问题。首先通过模拟仿真, 验证了方法的正确性;然后编写程序对已知移相量的实际干涉图进行计算。结果表明, 在干涉图对比度较好以及对计算的两点选择恰当的情况下, 计算出的相对误差在4%范围内。
Abstract
Wavelength-tuned interferometers are applied to test the large optical components. The phase-shifting calibration method based on the Lissajous figures technique is studied. Through the method, the phase-shifting steps between the interferograms are obtained. How to control the phase-shifting step to a certain value at different interferometric cavity lengths when testing is solved. The simulation program is written and it shows the method correct. The program is written to calculate the interferograms which the phase-shifting step is known. The result shows that the relative error is in the range of 4% when the contrast of the interferograms is good and the calculated points are selected correctly.

郭仁慧, 李建欣, 朱日宏, 陈磊. 基于Lissajous图技术的波长移相标定方法[J]. 光学技术, 2010, 36(2): 200. GUO Ren-hui, LI Jian-xin, ZHU Ri-hong, CHEN Lei. Wavelength-tuned phase-shifting calibration based on the Lissajous figures technique[J]. Optical Technique, 2010, 36(2): 200.

本文已被 2 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!