大气与环境光学学报, 2011, 6 (1): 47, 网络出版: 2011-02-22   

太阳光度计校正系数对气溶胶光学厚度测量误差的影响

Errors of Aerosol Optical Thichness Imposed by Calibration Coefficients of Sun-Photometer
作者单位
中国科学院安徽光学精密机械研究所 中国科学院大气成分与光学重点实验室, 安徽 合肥 230031
摘要
分析了太阳光度计测量大气气溶胶光学厚度时,由于校正系数的不确定性而产生的误差。利用推导的误差分 析公式,模拟了校正系数在不同相对误差的情况下导致的气溶胶光学厚度的绝对误差;模拟结果显示,校正 系数的相对误差小于+2%时可以满足气溶胶光学厚度的测量精度。
Abstract
The errors of atmospheric aerosol optical thickness obtained from Sun-photometer measurements, due to uncertainties of calibration coefficients, are analyzed. The precise formulae of errors analysis are deduced. The numerical experiments are carried out to investigate the absolute errors of aerosol optical thickness, caused by uncertainty of calibration coefficients. The simulative results indicate that calibration errors of less than +2% are suitable for aerosol optical thickness measurements.

徐青山, 伽丽丽, 周凯兵. 太阳光度计校正系数对气溶胶光学厚度测量误差的影响[J]. 大气与环境光学学报, 2011, 6(1): 47. XU Qing-shan, QIE Li-li, ZHOU Kai-bing. Errors of Aerosol Optical Thichness Imposed by Calibration Coefficients of Sun-Photometer[J]. Journal of Atmospheric and Environmental Optics, 2011, 6(1): 47.