强激光与粒子束, 2011, 23 (2): 403, 网络出版: 2011-03-02   

Z箍缩等离子体X射线凸晶谱仪

Convex crystal spectrometer for Z-pinch plasma X-ray diagnosis
作者单位
1 重庆工商大学 重庆市发展信息管理工程技术研究中心, 重庆 400067
2 重庆大学 光电技术及系统教育部重点实验室, 重庆 400030
3 中国工程物理研究院 流体物理研究中心, 四川 绵阳 621900
摘要
针对波长为0.3~0.5 nm的喷气箍缩等离子体X射线诊断, 研制了一种适用的高空间分辨的晶体谱仪。色散元件采用云母(002)凸面晶体, 布拉格角为37°, 信号采用X射线胶片进行接收, 有效接收面积为30 mm×80 mm。物理实验在“阳”加速器装置上进行, 胶片获得了氩喷气K, L壳层光谱信号, 其光谱范围较宽, 为0.31~0.40 nm。经解谱发现,类氦谱线有明显的基底, 用最小二乘法拟合包络曲线去噪处理后, 得到类氦谱线光谱分辨力为200~300。实验结果表明,该谱仪获得的X射线测量值与理论值相符, 适合喷气箍缩等离子体X射线光谱的诊断。
Abstract
To diagnose the gas-puff pinch plasma at wavelength range from 0.3 to 0.5 nm, a space-resolved curved crystal spectrometer has been developed based on Mica(002) crystal with Bragg angel of 37°. In experiments carried out on the Yang accelerator, X-ray spectra in an absolute intensity scale were obtained from argon gas-puff pinch plasmas, which were recorded by a 30 mm×80 mm flat film. The measured wavelength was consistent with the theoretical value, and the He-like spectral resolution was 200 to 300 for the mica crystal spectrometer, showing its potential in the analysis of X-ray emission from a gas-puff pinch plasmas source.

王洪建, 肖沙里, 施军, 黄显宾, 蔡红春, 钱家渝. Z箍缩等离子体X射线凸晶谱仪[J]. 强激光与粒子束, 2011, 23(2): 403. Wang Hongjian, Xiao Shali, Shi Jun, Huang Xianbin, Cai Hongchun, Qian Jiayu. Convex crystal spectrometer for Z-pinch plasma X-ray diagnosis[J]. High Power Laser and Particle Beams, 2011, 23(2): 403.

本文已被 3 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索