光学 精密工程, 2011, 19 (8): 1736, 网络出版: 2011-08-29   

凸面光栅成像光谱仪的干涉法装调

Interferometric alignment of imaging spectrometers with convex gratings
作者单位
1 中国科学院 长春光学精密机械与物理研究所, 吉林 长春 130033
2 中国科学院 研究生院, 北京 100049
摘要
研究了Offner凸面光栅成像光谱仪同心分光系统的装调方法。将双光束干涉基本原理与同心光学系统基本特点相结合, 提出了适用于Offner凸面光栅成像光谱仪的干涉装调法。该方法首先对凸面光栅成像光谱仪同心光学系统的两个凹球面反射镜进行同心装调, 由于单个球面反射镜的离焦量难以调整至零, 故引入相对离焦使两个反射镜的离焦量相等, 从而实现两镜同心。另外, 考虑无法用同一套凹球面补偿镜来检测凸球面, 故提出采用谱图直读法来确定凸面光栅位置, 实现了Offner凸面光栅成像光谱仪的高精度同心装调。实验结果表明, 经干涉法装调后, 分光系统主镜和三镜的球心距离可以达到几个纳米甚至更高。
Abstract
How to align the spectral systems of Offner concentric imaging spectrometers with convex gratings was researched. Based on the principle of dual-beam interference and characteristics of concentric systems,an interferometric alignment method for the Offner imaging spectrometers with convex gratings was proposed.With this method ,the primary and tertiary mirrors of Offner imaging spectrometers were aligned firstly. As the defocus of single spherical mirror could not be aligned to zero,the relative defocus was introduced to adjust the defocuses of the two mirrors to achieve the concentricity. Furthermore,because the concave compensator could not test the convex mirror,the convex grating was arranged according to reading spectrogram directly to complete concentric alignment with high precision. The result indicates that the interferometric alignment can achieve the dist

刘玉娟, 巴音贺希格, 崔继承, 唐玉国. 凸面光栅成像光谱仪的干涉法装调[J]. 光学 精密工程, 2011, 19(8): 1736. LIU Yu-juan, Bayanheshig, CUI Ji-cheng, TANG Yu-guo. Interferometric alignment of imaging spectrometers with convex gratings[J]. Optics and Precision Engineering, 2011, 19(8): 1736.

本文已被 4 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。