光学学报, 2012, 32 (4): 0430003, 网络出版: 2012-02-24   

近红外光谱无创生化检测中不定光程对模型精度的影响研究

Effect of Variable Optical Path Length on the Accuracy of the Model in Noninvasive Biochemical Detection by NIR Spectrum
作者单位
中国科学院长春光学精密机械与物理研究所应用光学国家重点实验室, 吉林 长春 130033
摘要
利用近红外血流容积光谱相减方法无创伤定量分析人体血液生化成分时,获得的在体血液光谱对应的样品光程是不确定的。为研究样品间的光程差异对定标模型预测精度的影响,配制了30份模拟血清样品,利用可变光程样品池,采用傅里叶光谱仪分别测量了相同光程和不定光程两组样品的近红外光谱。以分析血清白蛋白成分为例,对两组样品的定标模型精度进行比较,结果显示不定光程组的模型精度与相同光程组的模型精度相比明显下降, 交叉检验标准差(RMSECV)由110.0 mg/dL增大至156.0 mg/dL。采用多元散射校正算法对上述光谱数据校正后,RMSECV降低至98.1 mg/dL。对比分析处理前后两组模型的精度,证实了采用适当的预处理方法能够有效校正不定光程引起的光谱误差,提高模型预测精度。
Abstract
For noninvasive quantitative analysis of blood biochemical components, near infrared (NIR) subtracted blood volume spectrometry is applied. However, the optical path lengths corresponding to in vivo blood spectra obtained are uncertain. In order to study the effect of optical path length difference on the prediction accuracy of calibration model, 30 serum samples are simulated. A variable path-length cell is used to produce optical path difference, and spectra of two sample sets, with equal and different path-lengths respectively, are measured by Nicolet 6700 Fourier transform infrared (FT-IR) spectrometer. As to serum albumin, the calibration accuracies of the two groups are compared. The results show that the samples with uncertain optical path would reduce the calibration accuracy, as the root mean square error of cross validation (RMSECV) of the model increases from 110.0 mg/dL to 156.0 mg/dL. By applying multiplicative scatter correction, RMSECV of the uncertain optical path set reduces to 98.1 mg/L. It indicates that appropriate pretreatment method can effectively correct the error of spectrum caused by uncertain optical path length, thus improve the accuracy of the prediction model.

丁海泉, 卢启鹏, 陈星旦. 近红外光谱无创生化检测中不定光程对模型精度的影响研究[J]. 光学学报, 2012, 32(4): 0430003. Ding Haiquan, Lu Qipeng, Chen Xingdan. Effect of Variable Optical Path Length on the Accuracy of the Model in Noninvasive Biochemical Detection by NIR Spectrum[J]. Acta Optica Sinica, 2012, 32(4): 0430003.

本文已被 3 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!