光学学报, 2013, 33 (s2): s212005, 网络出版: 2013-12-25   

一种新型大孔径积分球辐射均匀性定标方法

Radiation Uniformity Calibration Method of Integrating Sphere with Large Aperture
作者单位
1 中国电子科技集团公司第四十一研究所, 山东 青岛 266555
2 电子测试技术重点实验室, 山东 青岛 266555
摘要
提出了基于参照探测器的积分球辐射均匀性定标方法并搭建了定标系统。对定标系统的定标精度进行了实验验证,结果表明:相对未考虑辐射漂移影响的情况,该系统对积分球辐射均匀性的绝对标定值提高了0.04%。对美国LabShpere公司USS2000型积分球的辐射均匀性分布曲线进行了标定,标定结果表明:在直径为20 cm的圆形扫描区域内,当探测器受光面与辐射孔径面间距D=0时,积分球辐射均匀性最好,达到99.8%;当D>80 cm时,均匀性基本达到饱和,接近99%;在D介于二者之间时积分球的辐射均匀性较差,不适合作为标准源。
Abstract
A method on illumination uniformity calibration of integrating sphere based on referential detector is proposed, and a calibration system is also built. Experiment has been done to validate the accuracy of the established calibration system. The result shows the absolute uniformity value has a 0.04% promotion. As an application, the USS2000 integrating sphere is calibrated with the calibration system, and the uniformity distribution law is depicted. It shows that, in the circular scanning area with the diameter of 20 μm, the illumination uniformity reaches 99.8% when D=0, where D is the distance between the aperture and the scan position, and the uniformity reaches 99% which is the saturation value when D>80 cm. However, when this distance locates in the region of 0

王恒飞, 吴斌, 刘红元, 应承平, 史学舜. 一种新型大孔径积分球辐射均匀性定标方法[J]. 光学学报, 2013, 33(s2): s212005. Wang Hengfei, Wu Bin, Liu Hongyuan, Ying Chengping, Shi Xueshun. Radiation Uniformity Calibration Method of Integrating Sphere with Large Aperture[J]. Acta Optica Sinica, 2013, 33(s2): s212005.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!