强激光与粒子束, 2014, 26 (1): 015104, 网络出版: 2014-01-17   

兰州重离子医疗装置同步环安装控制网设计

Survey network design of synchrotron in Heavy Ion Medical Machine in Lanzhou
作者单位
1 中国科学院 近代物理研究所, 兰州 730000
2 中国科学院大学, 北京 100039
摘要
介绍了兰州重离子医疗装置(HIMM)建造中准直安装方案里核心的测量控制网设计。同步环准直测量控制网是基于Leica AT401激光跟踪仪和综合测量软件SA建立的三维测量控制网,在控制网设计中采用9个基准柱和2个激光干涉测量得到的基准长度参考,以及Leica DNA03数字水准仪协同测量,使全局三维控制网设计精度达到0.04 mm,最终确保同步环四极磁铁安装精度达到0.10 mm的要求指标,为将来束流的稳定运行做好可靠的保证。
Abstract
The paper introduces control survey network of installation strategy in the new HIMM (Heavy Ion Medical Machine). The 3D survey network is based on laser tracker and SA (Spatial Analyzer). Nine fiducial references and two scale bars were designed to guarantee high accuracy in control survey network, and Digital Level was used for altitude. The final RMS error of the global network could reach 0.04 mm, which guarantees the transverse position of quadrupoles requirement (0.10 mm) on the synchrotron.

蔡国柱, 满开第, 张小奇, 陈文军, 王少明, 袁建东, 赵红卫. 兰州重离子医疗装置同步环安装控制网设计[J]. 强激光与粒子束, 2014, 26(1): 015104. Cai Guozhu, Man Kaidi, Zhang Xiaoqi, Chen Wenjun, Wang Shaoming, Yuan Jiandong, Zhao Hongwei. Survey network design of synchrotron in Heavy Ion Medical Machine in Lanzhou[J]. High Power Laser and Particle Beams, 2014, 26(1): 015104.