应用光学, 2014, 35 (1): 139, 网络出版: 2014-01-17   

单黑体温控的热成像仪MRTD野外在线自动检测技术研究

Field on-line automatic test technology of infrared thermal imager MRTD by single blackbody temperature control technology
作者单位
军械工程学院 军械技术研究所, 河北 石家庄 050000
摘要
为满足红外热像仪MRTD野外在线全自动检测要求,设计了一种新型MRTD检测方法。硬件采用单黑体温差控制技术,在减小检测设备体积质量的同时保证了黑体目标与环境背景的温差控制精度,采用串口通信控制与CCD图像采集处理等技术有机结合实现温差控制及图像采集处理的协调统一;在软件上采用优化的处理算法精确地提取目标及背景区域,并利用不同温差下目标与背景区域的灰度差计算出被测热像仪MRTD参数值。经实验检验,该结果与人眼观测所得结果误差在002℃以内,研制出的检测设备具有结构简单、体积及质量小、操作简便等优点。
Abstract
To meet online automatic infrared camera minimum resolvable temperature difference (MRTD) field testing requirements, a novel method for detecting MRTD was presented. In hardware aspect, the use of the single blackbody temperature control technology not only reduced the volume and quality of testing equipment but also guaranteed the accuracy of the temperature difference control between blackbody target and environment background, and by combining serial interface communication technology with CCD technology, the cooperation of the temperature control and image acquisition was achieved; In software aspect, the use of optimized algorithm could precisely extract the target and background region, and calculate the MRTD using gray difference between target and background region. The experiments show that the error is less than 002 ℃ compared with the results observed by eye. The testing equipment has simple structure, small volume, convenient operation, it can realize the automatic test of MRTD under field condition.

陈志斌, 宋岩, 张超, 刘羽翔, 侯章亚. 单黑体温控的热成像仪MRTD野外在线自动检测技术研究[J]. 应用光学, 2014, 35(1): 139. CHEN Zhi-bin, SONG Yan, ZHANG Chao, LIU Yu-xiang, HOU Zhang-ya. Field on-line automatic test technology of infrared thermal imager MRTD by single blackbody temperature control technology[J]. Journal of Applied Optics, 2014, 35(1): 139.

本文已被 2 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!