光学学报, 2014, 34 (7): 0712006, 网络出版: 2014-06-10   

大口径、宽动态范围红外测量系统辐射定标方法

Radiation Calibration Method for Infrared System with Large Aperture and Broad Dynamic Range
作者单位
1 中国科学院长春光学精密机械与物理研究所, 吉林 长春 130033
2 中国科学院大学, 北京 100049
摘要
现有的大口径、宽动态范围红外测量系统的辐射定标需要制备大口径红外平行光管,该方法机动性差且成本较高。为了解决这一问题,提出了一种基于内、外定标修正的辐射定标新方法。该方法通过切换反射镜将中、高温腔型黑体辐射引入红外光学系统,并对部分光路进行中、高温段的内定标;使用面源黑体对全系统进行中、低温段的外定标;提取公共温度范围的内、外定标数据并处理以获取内、外定标之间的修正系数;对中、高温段的内定标数据进行修正可以获取全系统在中、高温段的辐射定标数据,结合外定标结果就得到了全系统、宽动态范围的辐射定标数据。使用该定标方法对某Φ400 mm口径的红外测量系统进行辐射定标,并根据定标结果反演黑体的辐射亮度和温度,辐射亮度反演的最大误差为1.35%,温度反演的最大误差为0.76 ℃,实验结果证明了该定标方法的有效性。
Abstract
The existing radiation calibration for infrared system with large aperture and broad dynamic range needs huge infrared collimator, but this method has poor motility and cost much. In order to settle this question, a new calibration method basing on the amendment of inner and outer calibration is put forward. This method set a middle-high temperature cavity blackbody in infrared system, with the help of switching reflector, the cavity blackbody radiation is imported into optics system for the inner calibration on parts of system at middle-high temperature range, and an extended source blackbody is needed for the outer calibration on the whole system at middle-low temperature range. The inner and outer calibration data at common temperature range are picked up to calculate the amendatory coefficient. In the end, the whole system′s radiation calibration data at the middle-high temperature range are caught by amending the inner calibration data. Combining the outer calibration data at the middle-low temperature range, the whole system′s radiation calibration data at broad dynamic range are gained. Carrying through this method on an Φ400 mm infrared system and calculating the luminance error and temperature error, the maximal luminance error is 1.35% and the maximal temperature error is 0.76 ℃, the results indicate that the proposed method is effective.

孙志远, 常松涛, 朱玮. 大口径、宽动态范围红外测量系统辐射定标方法[J]. 光学学报, 2014, 34(7): 0712006. Sun Zhiyuan, Chang Songtao, Zhu Wei. Radiation Calibration Method for Infrared System with Large Aperture and Broad Dynamic Range[J]. Acta Optica Sinica, 2014, 34(7): 0712006.

本文已被 14 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!