Chinese Optics Letters, 2014, 12 (12): 121203, Published Online: Dec. 1, 2014  

Calibration error analysis of electro-optical detection systems Download: 594次

Author Affiliations
1 State Key Laboratory of Precision Measuring Technology and Instrument, Tianjin University, Tianjin 300072, China
2 School of Engineering and Design, Brunel University, Uxbridge, Middlesex UB8 3PH, UK
3 MOEMS Education Ministry Key Laboratory, Tianjin University, Tianjin 300072, China
Abstract
We present detailed analysis of calibration process error for electro-optical detection systems, which can be simplified as the plane rotation around a non-orthogonal axis. By means of octonions it firstly proves that the plane rotation around a non-orthogonal axis can be decomposed into rotations around two perpendicular axes. The rotation is further divided into three steps, and the calibration error is hence discussed and obtained. The simulation and test results indicate that there are large calibration errors in calibration process. The pointing error can be effectively improved after separating error components, which provides a more accurate set data for further compensation.

Qijian Tang, Xiangjun Wang, Qingping Yang. Calibration error analysis of electro-optical detection systems[J]. Chinese Optics Letters, 2014, 12(12): 121203.