光子学报, 2016, 45 (9): 0911001, 网络出版: 2016-10-19   

内窥式扫频光源光学相干层析系统的探头设计

Design of the Probe of Swept Source Optical Coherence Tomography for Endoscopic Imaging
作者单位
南京理工大学 电子工程与光电技术学院, 南京 210094
摘要
利用自聚焦透镜和单模光纤搭建了基于光纤旋转连接器的内窥式扫频光源光学相干层析成像系统.利用Zemax模拟了内窥探头的光学性能, 分析了光纤到自聚焦透镜的距离和自聚焦透镜节距对探头性能参量的影响.在综合考虑元器件成本以及成像要求后, 选择无需定制, 节距为0.24 Pitch的自聚焦透镜, 确定光纤与自聚焦透镜之间的距离为0.7 mm.经实验测得该系统的横向分辨率约为19.5 μm, 纵向分辨率约为9 μm, 工作距离约为7 mm, 成像深度为6.2 mm(空气中), 与理论值接近.为了验证该系统对生物组织的成像性能, 利用该系统对猪食道进行离体成像, 重建后的层析图中可以明显地观察到猪肠道的表皮层和固有层.测量系统的各项成像性能参量以及对生物组织的成像性能表明该探头在内窥成像中是可行的.
Abstract
A swept source endoscopic optical coherence tomography system was designed with the grin-lens, single mode fiber and fiber optic rotary joint for endoscopic imaging. The effects of the pitch of the grin lens and the distance between the single mode fiber and the grin-lens on system performance were then analyzed with Zemax. The object distance of 0.7 mm was chosen according to the diameter of esophagus and the Pitch of 0.24P was chosen according to the cost and the effects analyzed above. The measurement results shown that the axial resolution of the system is ~9 μm and the lateral resolution is ~19.5 μm with the imaging depth of 6 mm in air. The working distance of the probe is about 7 mm. The initial cross-sectional images of the intestinal tract of pig were presented to demonstrate the capability of the system for endoscopic applications. And all these results demonstrated that the probe designed is suitable for endoscopic imaging of esophagus.

卞海溢, 高万荣, 廖九零. 内窥式扫频光源光学相干层析系统的探头设计[J]. 光子学报, 2016, 45(9): 0911001. BIAN Hai-yi, Gao Wan-rong, LIAO Jiu-ling. Design of the Probe of Swept Source Optical Coherence Tomography for Endoscopic Imaging[J]. ACTA PHOTONICA SINICA, 2016, 45(9): 0911001.

本文已被 3 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote