大气与环境光学学报, 2016, 11 (5): 383, 网络出版: 2016-10-20   

激光诱导击穿光谱最佳探测时间的分析

Analysis on Optimal Detection Time of Laser-Induced Breakdown Spectroscopy
作者单位
中国科学院安徽光学精密机械研究所中国科学院环境光学与技术重点实验室, 安徽合肥 230031
摘要
为了确定激光诱导击穿光谱(laser-induced breakdown spectroscopy, LIBS)的最佳探测时间,采用时间分辨LIBS研究了样品中4种元素Fe、Pb、Ca和Mg的4条 特征谱线峰值强度以及相应的连续背景辐射光谱强度的衰减特性,对特征谱线峰值强度和连续背景辐射光谱强度的比值求极值,获得了特征谱线最佳信 背比(signal-to-background ratio, SBR)所对应时间的计算值。连续背景辐射和特征谱线光谱强度均呈指数形式衰减,4条特征谱线衰减时间分 别为1.42, 1.95, 1.69, 1.65 μs,相应的连续背景辐射衰减时间分别为0.85, 0.93, 0.89, 0.87 μs。 4条特征谱线最佳SBR所对应时间的计算值分别为1.43, 1.77, 1.62, 1.80 μs,最佳SBR所对应的时间的实验测 量值分别为1.5, 2.0, 1.5, 1.5 μs,二者具有较好的一致性。特征谱线峰值强度和连续背景辐射光谱强度的演化规 律能够对最佳SBR出现的时间作出预测。以上研究结果表明,LIBS最佳探测时间与连续背景辐射和特征谱线的衰减规律密切相关,通过对光谱强 度的比值求极值,可以预测谱线最佳SBR出现的时间,此方法为确定LIBS技术探测和分析的最佳时间提供了依据。
Abstract
In order to determine the optimal detection time of laser-induced breakdown spectroscopy(LIBS), time-resolved LIBS technique has been adopted to study the decay characteristics of peak intensity and spectral intensity of four elements Fe, Pb, Ca and Mg contained in the sample. The extreme value of ratio of peak intensity of characteristic lines and spectral intensity of continuous radiation is calculated, and the calculation value of time corresponding to the optimal signal-to-background ratio (SBR) of characteristic lines is obtained. The peak intensity of characteristic lines and spectral intensity of continuous radiation both decay in exponential form, the decay time of four characteristic lines are 1.42, 1.95, 1.69, 1.65 μs, and the corresponding decay time of continuous radiation are 0.85, 0.93, 0.89, 0.87 μs respectively. The calculation value of corresponding time of optimal SBR of four characteristic lines are 1.43, 1.77, 1.62, 1.80 μs, and experimental measurement value of corresponding time of optimal SBR are 1.5, 2.0, 1.5, 1.5 μs. They have a good consistency. The corresponding time of optimal SBR can be predicted by evolution law of peak intensity of characteristic lines and spectral intensity of continuous radiation. The results indicate that the optimal detection time of LIBS is closely related to decay law of continuous radiation and characteristic lines, and the corresponding time of optimal SBR of spectral lines can be predicted by evaluating extreme value of the ratio of spectral intensity. The method provides the basis for selecting the optimal time of LIBS detection and analysis.

余洋, 赵南京, 方丽, 孟德硕, 谷艳红, 王园园, 贾尧, 马明俊, 刘建国, 刘文清. 激光诱导击穿光谱最佳探测时间的分析[J]. 大气与环境光学学报, 2016, 11(5): 383. YU Yang, ZHAO Nanjing, FANG Li, MENG Deshuo, GU Yanhong, WANG Yuanyuan, JIA Yao, MA Mingjun, LIU Jianguo, LIU Wenqing. Analysis on Optimal Detection Time of Laser-Induced Breakdown Spectroscopy[J]. Journal of Atmospheric and Environmental Optics, 2016, 11(5): 383.

本文已被 2 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!