光学学报, 2018, 38 (5): 0512001, 网络出版: 2018-07-10   

角度受限的编码器误差测试研究 下载: 874次

Research on Encoder Error Test with Limited Angle
作者单位
中国科学院西安光学精密机械研究所, 陕西 西安 710119
摘要
为解决转角受限编码器角度误差的精确测试问题,采用平面反射镜-自准直经纬仪法测试编码器角度误差。建立了角度误差和位姿失调参数、编码器角度以及经纬仪示值之间关系的数学模型,通过对失调参数的解算可对编码器角度系统误差进行修正。实验结果表明,在角度范围为0°~40°时,由位姿失调所引入的角度误差随编码器角度的增大单调递增,误差最大值为742.9″;经过修正后的角度误差和编码器位姿无失调时的角度误差基本相当,误差最大值分别为4.4″和3.5″。此方法可有效测试无法精确调整或不具备调平条件的转角受限的编码器角度误差。
Abstract
In order to solve the accurate test problem of encoder angle error with limited angle , the angle error of encoder is tested by plane mirror-autocollimation theodolite. The mathematical model of the relationship between the angle error, the pose dislocation parameters, the angle of encoder and the value of the theodolite is established, the system error of the encoder angle can be corrected by calculating the pose dislocation parameters. The experimental results show that the angle error introduced by the pose dislocation increases monotonically with the increase of encoder angle when the angle range is 0°~40°, the maximum angle error is 742.9". The corrected angle error is basically the same as the angle error without encoder pose dislocation, the maximum angle errors are 4.4" and 3.5", respectively. Encoder angle error with limited angle that cannot be accurately adjusted or does not have a leveling condition can be effectively evaluated by this method.

王涛, 田留德, 赵建科, 周艳, 陈琛, 赵怀学, 刘锴, 潘亮, 万伟, 刘艺宁, 张婷. 角度受限的编码器误差测试研究[J]. 光学学报, 2018, 38(5): 0512001. Tao Wang, Liude Tian, Jianke Zhao, Yan Zhou, Chen Chen, Huaixue Zhao, Kai Liu, Liang Pan, Wei Wan, Yining Liu, Ting Zhang. Research on Encoder Error Test with Limited Angle[J]. Acta Optica Sinica, 2018, 38(5): 0512001.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!