光学技术, 2018, 44 (3): 269, 网络出版: 2018-06-09   

基于光学外差干涉的折射率测量方法

A refractive index measurement method based on optical heterodyne interferometry
作者单位
上海理工大学 医疗器械与食品学院 生物医学光学与视光学研究所, 上海 200093
摘要
基于光学折射率是材料的一个重要的物理参数, 提出了一种在泰曼格林干涉系统基础上的光学外差干涉折射率测量方法及系统。测量系统将样品放置在系统外, 采用双光束干涉探测, 使系统更加稳定、调整简单、测量更加精准。实验上测量了K9玻璃块样品在超连续激光光源中心波长790nm、带宽20nm下的折射率, 验证了方法的有效性和可靠性。实验结果表明, 系统能够准确测量出样品的折射率, 测量稳定性较高, 精度可以达到10-4; 适用于可见光和近红外波段透明物质的折射率测量, 应用领域广泛。
Abstract
The optical refractive index of a kind of material is an important physical parameter. A measurement method of the refractive index based on optical heterodyne interferometry using Twyman-Green interference system is proposed. The sample is placed outside the measurement system and detected by the double beam interferometry, which makes the system more stable, more accurate and easy to adjust. The refractive index of K9 glass block samples is measured experimentally by supercontinuum laser with 790nm center wavelength and 20nm bandwidth, in the process of which the validity and reliability of this method was verified. The experiment results indicate that this system is able to measure refractive indices with high stability and precision of 10-4. The system can be applicable to all refractive index measurement of transparent materials in the wavelength ranging from vision light to near infrared light.

蔡干, 王成, 董肖娜, 项华中, 郑刚, 陈明惠. 基于光学外差干涉的折射率测量方法[J]. 光学技术, 2018, 44(3): 269. CAI Gan, WANG Cheng, DONG Xiaona, XIANG Huazhong, ZHENG Gang, CHENG Minghui. A refractive index measurement method based on optical heterodyne interferometry[J]. Optical Technique, 2018, 44(3): 269.

本文已被 2 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!