激光技术, 2019, 43 (4): 527, 网络出版: 2019-07-10   

基于Talbot-Moiré效应长焦测焦仪的设计及校准技术研究

Calibration and design of long focal length measurement instruments based on Talbot-Moiré effect
作者单位
1 西南技术物理研究所, 成都 610041
2 陆军装备部航空军事代表局驻成都地区军事代表室, 成都 610036
摘要
为了满足100m~3km特长焦距参量的测试校准需求, 采用Talbot-Moiré条纹技术法, 设计了一套长焦测焦仪。针对影响仪器测量不确定度的系统参量, 分别设计了校准方法, 并对仪器测量不确定度评定结果进行了实验比对验证。结果表明, 校准后仪器测量不确定度评定为6.4%; 校准后的长焦测焦仪测量结果与刀口仪测量结果满足测量不确定度比对公式; 对比结果最大为0.64, 小于1; 长焦测焦仪测量不确定度评定合理。该研究验证了长焦测焦仪系统参量校准方法的可行性。
Abstract
In order to meet the test and calibration requirements of 100m~3km extra-long focal length, a set of long focal length measurement instrument was designed based on Talbot-Moiré fringes. Aiming at the system parameters affecting the measurement uncertainty, different calibration methods were designed respectively. Experiments were carried out to verify the results. The results show that, after calibration, the measurement uncertainty is 6.4%. Measurement results of the long focal length measurement instrument after calibration, and knife-edge instrument meet the measurement uncertainty formula. The maximum comparison result is 0.64, less than 1. Uncertainty evaluation of the long focal length measurement instrument is reasonable. The research validates the feasibility of calibration methods for system parameters of a long focal length measurement instrument.

张浩, 代红, 樊红英, 陈好, 蒋泽伟, 贾静, 孟庆安, 胡绍云. 基于Talbot-Moiré效应长焦测焦仪的设计及校准技术研究[J]. 激光技术, 2019, 43(4): 527. ZHANG Hao, DAI Hong, FAN Hongying, CHEN Hao, JIANG Zewei, JIA Jing, MENG Qing’an, HU Shaoyun. Calibration and design of long focal length measurement instruments based on Talbot-Moiré effect[J]. Laser Technology, 2019, 43(4): 527.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!