摘要

利用基于超连续激光和单色仪(SCM)的细分光谱扫描定标装置,对传感器的绝对光谱辐亮度响应度进行定标。利用两种绝对功率响应度溯源于低温绝对辐射计的辐亮度探测器(Trap-A 和Trap-B),分别以通用的部件级定标方式和基于该定标装置的系统级定标方式(以Trap-A作为参考),确定了Trap-B的绝对光谱辐亮度响应度,两种方式下的定标不确定度分别优于0.46%和1.8%。两种定标结果具有较好的一致性,在450~900 nm波段范围内,相对差异小于0.9%。研究结果表明:基于SCM的细分光谱扫描定标装置适用于传感器的绝对光谱辐射定标,在遥感器的绝对光谱辐射定标方面具有重要的应用价值。

Abstract

A spectrally resolved scanning calibration facility based on a supercontinuum laser and monochromator (SCM) was proposed for the calibration of the absolute spectral radiance responsivity of the sensors. Two radiance meters, labeled Traps-A and Traps-B, were used herein. The absolute spectral power responsivities of the meters were traced to a cryogenic absolute radiometer. The absolute spectral radiance responsivity of Trap-B was obtained by a general unit-level calibration method, and a system-level calibration method based on the SCM with Trap-A taken as the reference. The calibration uncertainties of Traps-A and Traps-B were lower than 0.46% and 1.8%, respectively. The calibration results by the two methods were in good agreement, with a relative difference of less than 0.9% in the 450-900 nm. The results imply that a spectrally-resolved scanning calibration facility based on SCM is suitable for the absolute spectral radiometric calibration of sensors and highly applicable to the absolute spectral radiometric calibration of remote sensors.

广告组5 - 光束分析仪
补充资料

中图分类号:O432.2

DOI:10.3788/AOS201939.0714002

所属栏目:激光器与激光光学

基金项目:国家自然科学基金;

收稿日期:2018-12-20

修改稿日期:2019-03-25

网络出版日期:2019-07-01

作者单位    点击查看

谢臣瑜:中国科学院安徽光学精密机械研究所通用光学定标与表征技术重点实验室, 安徽 合肥 230031中国科学技术大学, 安徽 合肥 230026
翟文超:中国科学院安徽光学精密机械研究所通用光学定标与表征技术重点实验室, 安徽 合肥 230031
李健军:中国科学院安徽光学精密机械研究所通用光学定标与表征技术重点实验室, 安徽 合肥 230031
高放:中国科学院安徽光学精密机械研究所通用光学定标与表征技术重点实验室, 安徽 合肥 230031中国科学技术大学, 安徽 合肥 230026
李越:中国科学院安徽光学精密机械研究所通用光学定标与表征技术重点实验室, 安徽 合肥 230031中国科学技术大学, 安徽 合肥 230026
吴浩宇:中国科学院安徽光学精密机械研究所通用光学定标与表征技术重点实验室, 安徽 合肥 230031
郑小兵:中国科学院安徽光学精密机械研究所通用光学定标与表征技术重点实验室, 安徽 合肥 230031

联系人作者:翟文超(wczhai@aiofm.ac.cn); 郑小兵(xbzheng@aiofm.ac.cn);

备注:国家自然科学基金;

【1】Gentile T R, Houston J M, Hardis J E et al. National Institute of Standards and Technology high-accuracy cryogenic radiometer. Applied Optics. 35(7), 1056-1068(1996).

【2】Stock D, Hofer H, White M et al. Lowest uncertainty direct comparison of a mechanically-cooled and a helium-cooled cryogenic radiometer. Metrologia. 37(5), 437-439(2000).

【3】Meindl P, Klinkmüller E, Werner L et al. New UV spectral responsivity scale of the PTB based on a cryogenic radiometer and an argon plasma arc radiation source. Metrologia. 43(2), S72-S77(2006).

【4】Xu N, Lin Y D, Gan H Y et al. Spectral responsivity calibration of silicon photodetectors using monochromator-based cryogenic radiometer. Proceedings of SPIE. 10155, (2016).

【5】Brown S W, Eppeldauer G P and Lykke K R. Facility for spectral irradiance and radiance responsivity calibrations using uniform sources. Applied Optics. 45(32), 8218-8237(2006).

【6】Schuster M, Nevas S, Sperling A et al. Spectral calibration of radiometric detectors using tunable laser sources. Applied Optics. 51(12), 1950-1961(2012).

【7】Zeng J N, Yoon H W, Eppeldauer G P et al. Spectral irradiance responsivity calibration of InSb radiometers using the improved IR-SIRCUS at NIST. Proceedings of SPIE. 7453, (2009).

【8】Zhang Y N, Zheng X B, Li J J et al. Research on system level calibration method of multi-band filter radiometer. Acta Optica Sinica. 33(8), (2013).
张艳娜, 郑小兵, 李健军 等. 多波段滤光片辐亮度计的系统级定标方法研究. 光学学报. 33(8), (2013).

【9】Zhai W C, Li J J, Zheng X B et al. Research on method of calibrating sun channels of sun radiometers. Acta Optica Sinica. 32(4), (2012).
翟文超, 李健军, 郑小兵 等. 太阳辐射计直射通道实验室定标方法研究. 光学学报. 32(4), (2012).

【10】Podobedov V B, Eppeldauer G P, Hanssen L M et al. Calibration of spectral responsivity of IR detectors in the range from 0.6 μm to 24 μm. Proceedings of SPIE. 9819, (2016).

【11】Levick A P, Greenwell C L, Ireland J et al. Spectral radiance source based on supercontinuum laser and wavelength tunable bandpass filter: the spectrally tunable absolute irradiance and radiance source. Applied Optics. 53(16), 3508-3519(2014).

【12】Yamaguchi Y, Yamada Y and Ishii J. Supercontinuum source-based system for pre-launch calibration of the hyperspectral sensor. Proceedings of SPIE. 8533, (2012).

【13】Yoo Y S, Kim G J, Park S et al. Spectral responsivity calibration of the reference radiation thermometer at KRISS by using a super-continuum laser-based high-accuracy monochromatic source. Metrologia. 53(6), 1354-1364(2016).

【14】Wang X X and Li Z G. Design and performance test of Gershun tube spectral radiometer. Acta Optica Sinica. 37(6), (2017).
王晓旭, 李志刚. Gershun管光谱辐射计设计与性能检测. 光学学报. 37(6), (2017).

【15】General Administration of Quality Supervision, Inspection, Quarantine of the People's Republic of China, . (2012).
国家质量监督检验检疫总局. (2012).

【16】Ding L, Yuan Y L, Zheng X B et al. Verification method of spectral calibration accuracy for hyperspectral remote sensors. Journal of Applied Optics. 38(3), 463-468(2017).
丁蕾, 袁银麟, 郑小兵 等. 高光谱遥感器现场光谱定标精度验证方法研究. 应用光学. 38(3), 463-468(2017).

【17】Yamaguchi Y, Yamada Y and Ishii J. Supercontinuum-source-based facility for absolute calibration of radiation thermometers. International Journal of Thermophysics. 36(8), 1825-1833(2015).

引用该论文

Xie Chenyu,Zhai Wenchao,Li Jianjun,Gao Fang,Li Yue,Wu Haoyu,Zheng Xiaobing. Absolute Radiance Responsivity Calibration Based on Supercontinuum Laser and Monochromator[J]. Acta Optica Sinica, 2019, 39(7): 0714002

谢臣瑜,翟文超,李健军,高放,李越,吴浩宇,郑小兵. 基于超连续激光-单色仪的绝对辐亮度响应度定标[J]. 光学学报, 2019, 39(7): 0714002

被引情况

【1】钱立勇,吴德成,周晓军,钟刘军,魏巍,王文举,王英检,龚威,刘东. 高光谱成像激光雷达系统辐射定标和地物信息获取. 光学学报, 2020, 40(11): 1128001--1

您的浏览器不支持PDF插件,请使用最新的(Chrome/Fire Fox等)浏览器.或者您还可以点击此处下载该论文PDF