红外与激光工程, 2018, 47 (4): 0404002, 网络出版: 2018-09-19   

双通道热红外标准辐亮度计定标方法研究

Reseach on calibration method of double-channel thermal-infrared standard radiometer
作者单位
1 中国科学院安徽光学精密机械研究所 通用光学定标与表征技术重点实验室, 安徽 合肥 230031
2 中国科学技术大学, 安徽 合肥 230026
摘要
黑体辐射源作为初级标准, 其辐射特性的测量精度决定了整个标准传递链路的不确定度。为了提高黑体的辐亮度测量精度, 研制了一台可直接观测黑体辐亮度的双通道热红外标准辐亮度计。根据仪器的工作原理, 采用自行研制的高精度水浴黑体获取标准辐亮度计的定标系数。实验结果表明, 该辐亮度计的1 h信号非稳定性优于0.3%, 定标系数拟合不确定度优于1.23%, 5 μm通道整体不确定度为0.39%, 10 μm通道整体不确定度为1.3%。可实现黑体辐亮度与实验室辐射标准之间的传递。
Abstract
Blackbody radiation source as a primary standard, its measurement accuracy determines the uncertainty level of the whole traceability link. To improve the measurement accuracy of blackbodies′ radiance, a double-channel thermal-infrared standard radiometer was developed. According to the principle of the instrument, calibration coefficients of the standard radiometer were obtained by using a high accuracy water-bath blackbody. Experimental results suggest that the radiometer′s instability is less than 0.3% within one hour, uncertainty of fitting coefficient is less than 1.23%, and combined uncertainty is less than 0.39%, 1.3% in 5 μm and 10 μm channel respectively. This radiometer provides new means for tracing blackbody radiance directly to radiant standard in lab.
参考文献

[1] Xu Jun, Meng Binghuan, Zheng Xiaobing, et al. Calibration of common temperature blackbody based on thermal-infrared standard radiometer[J]. Infrared and Laser Engineering, 2013, 43(3): 716-721. (in Chinese)

[2] Chen Muwang, Qi Hongxing, Gong Yumei, et al. Analysis of effect of calibration blackbody temperature stability on the infrared imaging[J]. Infrared and Laser Engineering, 2006, 35(6): 652-654. (in Chinese)

[3] Rice J P, Johnson B C. The NIST EOS thermal infrared transfer radiometer[J]. Metrologia, 2010, 35: 505-509.

[4] Rice J P, Bender S C, Atkins W H, et al. Deployment test of the NIST EOS thermal-infrared transfer radiometer[J]. International Journal of Remote Sensing, 2003, 24(2): 367-388.

[5] Adriaan C C, Raju U D, Timothy M J, et al. Low-background temperature calibration of infrared blackbodies [J]. Metrologia, 2006, 43: 46-50.

[6] Yue Wenlong. Research of the calibration technology for blackbody radiation source[D]. Xi′an: Xi′an Technological University, 2007. (in Chinese)

[7] Zhu Qihai, Li Xiansheng, Ren Jianwei, et al. Development and calibration of infrared double-channel spectral radiometer[J]. Optical Technique, 2006, 32(5): 773-778. (in Chinese)

[8] Zhang Lei, Zheng Xiaobing, Zhang Liming, et al. High accuracy water bath blackbody based on the information quantification of infrared remote sensing[J]. Optical Technique, 2007, 33(2): 245-251. (in Chinese)

[9] Zhang Lei. Study on the spectral radiometric calibration technique of infrared stand transfer detector[D]. Hefei: Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, 2007. (in Chinese)

[10] Fang Qianqian, Fang Wei, Wang Kai, et al. Calculation of effective of blackbody cavities by Monte-Carlo method [J] Chinese Optics, 2012, 5(2): 167-173. (in Chinese)

[11] Xun Jun. Research on calibration of ambient temperature blackbodies based on a thermal-infrared standard Radiometer [D]. Hefei: Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, 2013. (in Chinese)

谢臣瑜, 李健军, 庞伟伟, 夏茂鹏, 郑小兵. 双通道热红外标准辐亮度计定标方法研究[J]. 红外与激光工程, 2018, 47(4): 0404002. Xie Chenyu, Li Jianjun, Pang Weiwei, Xia Maopeng, Zheng Xiaobing. Reseach on calibration method of double-channel thermal-infrared standard radiometer[J]. Infrared and Laser Engineering, 2018, 47(4): 0404002.

本文已被 2 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!