激光与光电子学进展, 2015, 52 (8): 081701, 网络出版: 2015-07-29   

组织荧光光谱检测系统校准方法研究 下载: 514次

Calibration for Tissue Fluorescence Spectroscopy System
作者单位
1 中国科学院合肥物质科学研究院, 安徽 合肥 230031
2 皖江新兴产业技术发展中心, 安徽 铜陵244000
摘要
组织荧光光谱技术可用来定量测量组织生物化学性质及组织结构的变化,进而实现疾病早期诊断.为提高组织荧光光谱检测的准确性,提出了组织荧光光谱检测系统校准流程及具体实现方法,包括激发光强校准、传输光路传光效率校准、探测模块波长校准和非线性校准、系统光谱响应校准、暗背景扣除、光谱平滑.搭建三套组织荧光光谱检测系统,并分别对荧光标准板和人体皮肤组织进行实验测试,结果显示未校准时三套系统对同一荧光板和同一受试者的荧光强度测量结果变异系数分别为7.4%和10.4%,校准后变异系数分别为0.9%和2.0%.
Abstract
Fluorescence spectroscopy provides a noninvasive method to quantify biochemical and structural changes in tissues and it is regarded as an early subtle indicator of disease or injury.One challenge facing the rapid growth of fluorescence application is the inability to insure the accuracy of fluorescence measurements.The calibration protocols that include excitation energy calibration,collection efficiency calibration,wavelength calibration,system nonlinearity calibration,system response calibration,background subtraction and spectrum signal filtering,for optical instruments that ensure measurement accuracy are proposed.Three fluorescence spectroscopy systems are built and the fluorescence spectra of fluorescence standard and human skin are collected.The result shows that the spectrum intensity variation coefficient among these systems for the same reference standard is 7.4% and for the same human skin is 10.4% before calibration.By comparison,the spectrum intensity variation coefficients after calibration are 0.9% and 2.0%,respectively.
参考文献

[1] Bradley R S,Thorniley M S.A review of attenuation correction techniques for tissue fluorescence[J].Journal of the Royal Society Interface,2006,3(6):1-13.

[2] 苏亚,孟卓,王龙志,等.光学相干层析无创血糖检测中相关性分析及标定[J].中国激光,2014,41(7):0704002.

    Su Ya,Meng Zhuo,Wang Longzhi,et al..Correlation analysis and calibration of noninvasive blood glucose monitoring in vivo with optical coherence tomography[J].Chinese J Lasers,2014,41(7):0704002.

[3] Kim A,Roy M,Dadani F,et al..A fiberoptic reflectance probe with multiple source- collector separations to increase the dynamic range of derived tissue optical absorption and scattering coefficients[J].Optics Express,2010,18(6):5580-5594.

[4] 李飞,张元志,王贻坤,等.基于实数遗传算法的混浊介质光学参数提取[J].光学学报,2013,33(12):1217001.

    Li Fei,Zhang Yuanzhi,Wang Yikun,et al..Determination of optical parameters of turbid media based on real coded genetic algorithm[J].Acta Optica Sinica,2013,33(12):1217001.

[5] Zeng H,Macaulay C,Mclean D I,et al..The dynamics of laser- induced changes in human skin autofluorescence—experimental measurements and theoretical modeling[J].Photochemistry Photobiology,1998,68(2):227-236.

[6] 殷高方,赵南京,胡丽,等.基于色素特征荧光光谱的浮游植物分类测量方法[J].光学学报,2014,34(9):0930005.

    Yin Gaofang,Zhao Nanjing,Hu Li,et al..Classified measurement of phytoplankton based on characteristic fluorescence of photosynthetic pigments[J].Acta Optica Sinica,2014,34(9):0930005.

[7] Mackinnon N,Vasefi F,Farkas D L.Toward in-vivo diagnosis of skin cancer using multimode imaging dermoscopy:(I)clinical system development and validation[C].SPIE,2014,8947:89470I.

[8] Zeng H,Korbelik M,Mclean D I,et al..Monitoring photoproduct formation and photobleaching by fluorescence spectroscopy has the potential to improve PDT dosimetry with a verteporfin- like photosensitizer[J].Photochemistry Photobiology,2002,75(4):398-405.

[9] Valdés P A,Leblond F,Kim A,et al..A spectrally constrained dual- band normalization technique for protoporphyrin IX quantification in fluorescence-guided surgery[J].Optics Letters,2012,37(11):1817-1819.

[10] Zhao J,Lui H,Mclean D I,et al..Toward instrument- independent quantitative measurement of fluorescence intensity in fiber-optic spectrometer systems[J].Applied Optics,2007,46(29):7132-7140.

[11] Marin N M,Marckinnon N,MacAulay C,et al..Calibration standards for multicenter clinical trials of fluorescence spectroscopy for in vivo diagnosis[J].Journal of Biomedical Optics,2006,11(1):014010.

[12] Zhang Y,Zhu L,Wang Y,et al..Classification of skin autofluorescence spectrum using support vector machine in type 2diabetes screening[J].Journal of Innovative Optical Health Sciences,2013,6(4):1350036.

[13] 刘红元,王恒飞,马书民,等.一种CCD器件响应度非均匀性校准装置研制[J].中国激光,2014,41(s1):s116003.

    Liu Hongyuan,Wang Hengfei,Ma Shumin,et al..A calibration equipment of responsivity non- uniformity for charge coupled devices[J].Chinese J Lasers,2014,41(s1):s116003.

[14] 李晓杰,任建伟,刘洪兴,等.面阵CCD 光谱响应测试及不确定度评估[J].激光与光电子学进展,2014,51(11):111202.

    Li Xiaojie,Ren Jianwei,Liu Hongxing,et al..Spectral response testing and uncertainty evaluation of plane array CCD [J].Laser & Optoelectronics Progress,2014,51(11):111202.

张元志, 潘亮, 王贻坤, 候华毅, 朱灵, 王安, 刘勇. 组织荧光光谱检测系统校准方法研究[J]. 激光与光电子学进展, 2015, 52(8): 081701. Zhang Yuanzhi, Pan Liang, Wang Yikun, Hou Huayi, Zhu Ling, Wang An, Liu Yong. Calibration for Tissue Fluorescence Spectroscopy System[J]. Laser & Optoelectronics Progress, 2015, 52(8): 081701.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!