傅里叶变换红外光谱仪动镜倾斜和动态校准研究

Tilt and Dynamic Alignment for the Moving Mirror in the Fourier Transform Infrared Spectrometer
作者单位
中国科学院安徽光学精密机械研究所 中国科学院环境光学与技术重点实验室,合肥 230031
摘要
分析了迈克尔逊干涉仪中动镜和定镜在圆形通光孔径情况下动镜在扫描过程中倾斜对干涉调制度的影响.为满足干涉调制度要求,理论计算发现动镜倾斜角度需控制在1″以内,对应的相位误差为±6.6°以内.为了解决此问题,使动镜和定镜保持准直,采用定镜动态校准系统对动镜倾斜误差进行校准.实验结果表明,校准后的XR和YR相位误差的均方根分别为1.02°和1.25°,满足系统要求.动态校准系统能够有效地校准动镜长距离运动过程中发生的倾斜误差,对高分辨光谱仪的设计和研制具有一定的指导意义.
Abstract
Theeffect of the moving mirror tilt errors on the modulation depth wasanalyzed for the circular aperture in Michelson interferometer. To meet the requirements of interference modulation, the tilt angle of the moving mirror is best controlled within 1″ under the theoretical calculation, corresponding to the phase error within ±6°. In order to solve the key problem, the dynamic alignment system wasused to align the tilt error of the moving mirror ,and make alignment between the moving mirror and the fixed mirror. The experimental results show that the root mean square of phase errors XR and YR is 1.02°and 1.25° respectively by dynamic alignment system to satisfy the systems requirements.It is shown that the dynamic system can effectively align the tilt error of the moving mirror in the longer distance movement process. It is a certain guiding significance to design and develop for the spectrometer with higher resolution.

冯明春, 徐亮, 金岭, 刘文清, 高闽光, 李胜, 李相贤, 童晶晶, 刘建国. 傅里叶变换红外光谱仪动镜倾斜和动态校准研究[J]. 光子学报, 2016, 45(4): 0412005. FENG Ming-chun, XU Liang, JIN Ling, LIU Wen-qing, GAO Min-guang, LI Sheng, LI Xiang-xian, TONG Jing-jing, LIU Jian-guo. Tilt and Dynamic Alignment for the Moving Mirror in the Fourier Transform Infrared Spectrometer[J]. ACTA PHOTONICA SINICA, 2016, 45(4): 0412005.

本文已被 4 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!