首页 > 论文 > Chinese Optics Letters > 15卷 > 5期(p:051701)

Experimental assessment of a 3-D plenoptic endoscopic imaging system

  • 摘要
  • 论文信息
  • 参考文献
  • 被引情况
  • PDF全文
分享:

Abstract

An endoscopic imaging system using a plenoptic technique to reconstruct 3-D information is demonstrated and analyzed in this Letter. The proposed setup integrates a clinical surgical endoscope with a plenoptic camera to achieve a depth accuracy error of about 1 mm and a precision error of about 2 mm, within a 25 mm×25 mm field of view, operating at 11 frames per second.

Newport宣传-MKS新实验室计划
补充资料

DOI:10.3788/col201715.051701

所属栏目:Medical optics and biotechnology

收稿日期:2016-10-10

录用日期:2017-01-24

网络出版日期:2017-03-01

作者单位    点击查看

Hanh N. D. Le:Electrical and Computer Engineering Department, Johns Hopkins University, Baltimore, MD 21 21 8, USA
Ryan Decker:Sheikh Zayed Institute for Pediatric Surgical Innovation, Children’s National Health System, Washington, DC 2 0010, USA
Axel Krieger:Sheikh Zayed Institute for Pediatric Surgical Innovation, Children’s National Health System, Washington, DC 2 0010, USA
and Jin U. Kang:Electrical and Computer Engineering Department, Johns Hopkins University, Baltimore, MD 21 21 8, USA

联系人作者:联系作者(hle18@jhu.edu)

备注:This work was supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under Grant no. 1R01EB020610. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. We thank Justin Opfermann and Dr. Xinyang Liu (Children’s National Health System) for the use of their surgical endoscope.

【1】P. Bucher, F. Pugin, S. Ostermann, F. Ris, M. Chilcott, and P. Morel, Surg. Endosc. 25, 408 (2011).

【2】C. M. Kang, D. H. Kim, W. J. Lee, and H. S. Chi, Surg. Endosc. 25, 2004 (2011).

【3】R. Smith, A. Day, T. Rockall, K. Ballard, M. Bailey, and I. Jourdan, Surg. Endosc. 26, 1522 (2012).

【4】A. Shademan, R. S. Decker, J. D. Opfermann, S. Leonard, A. Krieger, and P. C. Kim, Sci. Transl. Med. 8, 337ra64 (2016).

【5】S. Rohl, S. Bodenstedt, S. Suwelack, H. Kenngott, B. P. Müller-Stich, R. Dillmann, and S. Speidel, Med. Phys. 39, 1632 (2012).

【6】N. Atzpadin, P. Kauff, and O. Schreer, IEEE Trans. Circuits Syst. Video Technol. 14, 321 (2004).

【7】D. Stoyanov, M. V. Scarzanella, P. Pratt, and G. Z. Yang, Med. Image Comput. Comput. Assist. Intervention 6361, 275 (2010).

【8】D. Stoyanov, Med. Image Comput. Comput. Assist. Interv. 7510, 479 (2012).

【9】P. L. Chang, D. Stoyanov, A. Davison, and P. E. Edwards, Med. Image Comput. Comput. Assist. Interv. 8149, 42 (2013).

【10】J. Penne, K. H?ller, M. Stürmer, T. Schrauder, A. Schneider, R. Engelbrecht, H. Feu?ner, B. Schmauss, and J. Hornegger, Med. Image Comput. Comput. Assist. Interv. 5761, 467 (2009).

【11】A. Groch, A. Seitel, S. Hempel, S. Speidel, R. Engelbrecht, J. Penne, K. Holler, S. Rohl, K. Yung, S. Bodenstedt, and F. Pflaum, Proc. SPIE 7964, 796415 (2011).

【12】P. Fursattel, S. Placht, C. Schaller, M. Balda, H. Hofmann, A. Maier, and C. Riess, IEEE Trans. Comput. Imaging 2, 27 (2016).

【13】L. Maier-Hein, P. Mountney, A. Bartoli, H. Elhawary, D. Elson, A. Groch, A. Kolb, M. Rodrigues, J. Sorger, S. Speidel, and D. Stoyanov, Med. Image Anal. 17, 974 (2013).

【14】S. Zhang, Opt. Lasers Eng. 48, 149 (2010).

【15】H. Du, and Z. Wang, Opt. Lett. 32, 2438 (2007).

【16】T. T. Wu, and J. Y. Qu, Opt. Express 15, 10421 (2007).

【17】C. Schmalz, F. Forster, A. Schick, and E. Angelopoulou, Med. Image Anal. 16, 1063 (2012).

【18】J. Yagnik, G. S. Gorthi, K. R. Ramakrishnan, and L. K. Rao, in TENCON 2005–2005 IEEE Region 10 Conference (2005), p. 1.

【19】T. Hain, R. Eckhardt, K. Kunzi-Rapp, and B. Schmitz, Med. Laser Appl. 17, 55 (2002).

【20】P. F. Gomes, M. Sesselmann, C. D. Faria, P. A. Araújo, and L. F. Teixeira-Salmela, J. Biomech. 43, 1215 (2010).

【21】N. T. Clancy, D. Stoyanov, L. Maier-Hein, A. Groch, G. Z. Yang, and D. S. Elson, Biomed. Opt. Express 2, 3119 (2011).

【22】National Instruments, “3D Imaging with NI LabVIEW,” http://www.ni.com/white-paper/14103/en/ (2016).

【23】R. Rao, “Stereo and 3D vision ,” https://courses.cs.washington.edu/courses/cse455/09wi/Lects/lect16.pdf (2009).

【24】H. Nguyen, Z. Wang, and J. Quisberth, Adv. Opt. Methods Exp. Mech. 3, 195 (2016).

【25】T. Georgiev, and A. Lumsdaine, Proc. SPIE 8299, 829908 (2012).

【26】C. Perwass, and L. Wietzke, Proc. SPIE 8291, 829108 (2012).

【27】R. Decker, A. Shademan, J. Opfermann, S. Leonard, P. C. Kim, and A. Krieger, Proc. SPIE 9494, 94940B (2015).

【28】A. Shademan, R. S. Decker, J. Opfermann, S. Leonard, P. C. Kim, and A. Krieger, in IEEE International Conference on Robotics and Automation (ICRA) (2016), p. 708.

【29】C. Heinze, S. Spyropoulos, S. Hussmann, and C. Perwass, IEEE Trans. Instrum. Meas. 65, 1197 (2016).

【30】O. Johannsen, C. Heinze, B. Goldluecke, and C. Perwa?, “On the calibration of focused plenoptic cameras,” in Time-of-Flight and Depth Imaging. Sensors, Algorithms, and Applications (Springer, 2013), p.?302.

您的浏览器不支持PDF插件,请使用最新的(Chrome/Fire Fox等)浏览器.或者您还可以点击此处下载该论文PDF