Time constant optimization of solar irradiance absolute radiometer

Author Affiliations
1 Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
2 University of Chinese Academy of Sciences, Beijing 100049, China
3 School of Mechanical Engineering, University of South China, Hengyang 421001, China
Abstract
We experimentally evaluate and optimize the time constant of solar irradiance absolute radiometer (SIAR). The systemic error introduced by variable time constant is studied by a finite element method. The results shown that, with a classic time constant of 30 s for SIAR, the systemic errors are 0.06% in the midday and 0.275% in the morning and afternoon. The uncertainty level which can be considered negligible for SIAR is also investigated, and it is suggested that the uncertainty level has to be less than 0.02%. Then, combining the requirement of international comparison with these two conclusions, we conclude that the suitable time constant for SIAR is 20 s.

TANG Xiao, FANG Wei, WANG Yu-peng, YANG Dong-jun, YI Xiao-long. Time constant optimization of solar irradiance absolute radiometer[J]. 光电子快报(英文版), 2017, 13(3): 179.