光电子快报(英文版), 2017, 13 (3): 179, Published Online: Sep. 13, 2018
Time constant optimization of solar irradiance absolute radiometer
Abstract
We experimentally evaluate and optimize the time constant of solar irradiance absolute radiometer (SIAR). The systemic error introduced by variable time constant is studied by a finite element method. The results shown that, with a classic time constant of 30 s for SIAR, the systemic errors are 0.06% in the midday and 0.275% in the morning and afternoon. The uncertainty level which can be considered negligible for SIAR is also investigated, and it is suggested that the uncertainty level has to be less than 0.02%. Then, combining the requirement of international comparison with these two conclusions, we conclude that the suitable time constant for SIAR is 20 s.
TANG Xiao, FANG Wei, WANG Yu-peng, YANG Dong-jun, YI Xiao-long. Time constant optimization of solar irradiance absolute radiometer[J]. 光电子快报(英文版), 2017, 13(3): 179.