中国激光, 2018, 45 (11): 1107001, 网络出版: 2018-11-15   

基于光声相关谱法的血液流速测量 下载: 944次

Measuring Velocity of Blood Flow Based on Photoacoustic Correlation Spectroscopy
作者单位
1 中国科学院上海光学精密机械研究所信息光学与光电技术实验室, 上海 201800
2 中国科学院大学, 北京 100049
3 江南大学理学院, 江苏 无锡 214122
摘要
利用光声相关谱法对血液流速进行了测量,并研究了激光重复频率以及血流方向与入射激光传播方向的夹角对血液流速测量准确性的影响。结果表明:血流速度越快,要求激光的重复频率越高;当血流方向与入射激光的传播方向垂直时,系统可以测量的血液流速范围为0.059~92.3 mm/s,测量值与真实值的相关系数为0.992;当血流方向与入射激光的传播方向不垂直时,测得的血液流速和实际血液流速的比值与样品倾斜角度成余弦关系。
Abstract
Velocity of blood flow is measured by photoacoustic correlation spectroscopy. The effects of the laser repetition frequency and the angle between the blood flow direction and incidence laser propagation direction on the measurement of velocity of blood flow are studied. Research results show that the faster the blood flow velocity, the higher the laser repetition frequency required. When the blood flow orientation is perpendicular to the direction of incident laser propagation direction, the range of velocity of blood flow that the system can measure is 0.059-92.3 mm/s. The correlation coefficient between the measured flow velocities and the actual flow velocities is 0.992. When the blood flow orientation is not perpendicular to the direction of incident laser propagation direction, the ratios of the measured velocities of blood flow to the original ones are cosine to the tilted angle of sample.

潘柳华, 张向阳, 李中梁, 南楠, 步扬, 陈艳, 王瑄, 王向朝. 基于光声相关谱法的血液流速测量[J]. 中国激光, 2018, 45(11): 1107001. Pan Liuhua, Zhang Xiangyang, Li Zhongliang, Nan Nan, Bu Yang, Chen Yan, Wang Xuan, Wang Xiangzhao. Measuring Velocity of Blood Flow Based on Photoacoustic Correlation Spectroscopy[J]. Chinese Journal of Lasers, 2018, 45(11): 1107001.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!