水雾对调Q Er:YAG激光消融牙本质阈值及损伤的影响

Effect of Water Mist on Threshold and Damage of Dentin Ablation with Q-Switched Er:YAG Laser
作者单位
1 中国科学院合肥物质科学研究院医学物理与技术中心, 安徽 合肥 230031
2 中国科学技术大学, 安徽 合肥 230026
3 中国科学院合肥物质科学研究院肿瘤医院, 安徽 合肥 230031
摘要
研究了调Q Er:YAG激光消融离体牙本质过程中水雾的作用及其对消融后牙本质微结构的影响。调Q Er:YAG激光重复频率为3 Hz,脉冲宽度为62 ns,作用脉冲数为10,能量范围为4~70 mJ,水雾流量分别为0,4,9,13 mL/min时消融牙本质。每种实验条件下消融10个牙本质样品,利用扫描电镜观察消融坑洞微结构。实验结果表明,随着水雾流量的增加,调Q Er:YAG激光对人牙本质的消融阈值呈逐渐增加的趋势,牙本质所受机械损伤减小。实验结果对了解调Q Er:YAG激光消融牙硬组织的微观结构及牙科治疗应用具有参考价值。
Abstract
The effect of water mist on the extracted dentin ablation by Q-switched Er:YAG laser is investigated, and the influence of water mist on the ablated dentin microstructure is studied. The Q-switched Er:YAG laser with 62 ns pulse width is used to ablate the human dentin at the repetition rate of 3 Hz, the number of pulses is 10, and the energy range is from 4 mJ to 70 mJ. Different flow rates of water mist (0, 4, 9, 13 mL/min) are applied during the experiments. 10 dentin samples are ablated under each experimental condition. A scanning electron microscope is used to observe the microstructures of ablated holes. The experiment result shows that the increase of flow rate of water mist makes the Q-switched Er:YAG laser ablation threshold to dentin rise, and the dentin is protected from mechanical damage. The results are guided for acquiring microstructures during dentin ablation with Q-switched Er:YAG laser and practical treatment in dentistry.
参考文献

[1] Yang J W, Wang L, Wu X Y, et al. High peak power Q-switched Er:YAG laser with two polarizers and its ablation performance for hard dental tissues[J]. Optics Express, 2014, 22(13): 15686-15696.

[2] Hibst R, Keller U. Experimental studies of the application of the Er:YAG laser on dental hard substances: I. Measurement of the ablation rate[J]. Lasers in Surgery and Medicine, 1989, 9(4): 338-344.

[3] Pelagalli J, Gimbel C B, Hansen R T, et al. Investigational study of the use of Er:YAG laser versus dental drill for caries removal and cavity preparation-phase I[J]. Journal of Clinical Laser Medicine & Surgery, 1997, 15(3): 109-115.

[4] Hibst R. Lasers for caries removal and cavity preparation: state of the art and future directions[J]. Journal of Oral Laser Applications, 2002, 2(4): 203-212.

[5] Aoki A, Ishikawa I, Yamada T, et al. Comparison between Er:YAG laser and conventional technique for root caries treatment in vitro[J]. Journal of Dental Research, 1998, 77(6): 1404-1414.

[6] Mehl A, Folwaczny M, Haffner C, et al. Bactericidal effects of 2.94 μm Er:YAG-laser radiation in dental root canals[J]. Journal of Endodontics, 1999, 25(7): 490-493.

[7] Walsh J T, Flotte T J, Deutsch T F. Er:YAG laser ablation of tissue: effect of pulse duration and tissue type on thermal damage[J]. Lasers in Surgery and Medicine, 1989, 9(4): 314-326.

[8] Papagiakoumou E, Papadopoulos D N, Khabbaz M G, et al. The influence of the Q-switched and free-running Er:YAG laser beam characteristics on the ablation of root canal dentine[J]. Applied Surface Science, 2004, 233(1): 234-243.

[9] Jelínková H, Dostalova T, Němec M, et al. Free-running and Q-switched Er:YAG laser dental cavity and composite resin restoration[J]. Laser Physics Letters, 2007, 4(11): 835-839.

[10] Fried D, Shori R K, Duhn C W. Back spallation due to ablative recoil generated during Q-switched Er:YAG ablation of dental hard tissue[C]. SPIE, 1998, 3248: 78-85.

[11] 林实, 彭奇明, 詹振林, 等. Er:YAG 激光和 Er, Cr\:YSGG 激光消融牙釉质的阈值研究[J]. 中国激光, 2009, 36(10): 2598-2602.

    Lin Shi, Peng Qiming, Zhan Zhenlin, et al. Study of the ablation threshold of Er:YAG and Er, Cr\:YSGG lased dental enamel[J]. Chinese J Lasers, 2009, 36(10): 2598-2602.

[12] Apel C, Franzen R, Meister J, et al. Influence of the pulse duration of an Er:YAG laser system on the ablation threshold of dental enamel[J]. Lasers in Medical Science, 2002, 17(4): 253-257.

[13] Apel C, Meister J, Ioana R S, et al. The ablation threshold of Er:YAG and Er\:YSGG laser radiation in dental enamel[J]. Lasers in Medical Science, 2002, 17(4): 246-252.

[14] Hossain M, Nakamura Y, Yamada Y, et al. Ablation depths and morphological changes in human enamel and dentin after Er:YAG laser irradiation with or without water mist[J]. Journal of Clinical Laser Medicine & Surgery, 1999, 17(3): 105-109.

崔庆哲, 魏蒙恩, 凌琳, 程庭清, 吴先友, 江海河. 水雾对调Q Er:YAG激光消融牙本质阈值及损伤的影响[J]. 中国激光, 2016, 43(12): 1201003. Cui Qingzhe, Wei Meng′en, Ling Lin, Cheng Tingqing, Wu Xianyou, Jiang Haihe. Effect of Water Mist on Threshold and Damage of Dentin Ablation with Q-Switched Er:YAG Laser[J]. Chinese Journal of Lasers, 2016, 43(12): 1201003.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!