应用光学, 2019, 40 (3): 478, 网络出版: 2019-06-10   

近紫外到近红外光谱辐射计及定标方法研究

Research on near-UV to near-IR spectroradiometer and its calibration method
作者单位
1 西安应用光学研究所,陕西 西安 710065
2 陆军装备部航空军代局驻西安地区航空军代室,陕西 西安 710065
摘要
为实现目标光谱辐射亮度的高精度测量,研制了一种小视场近紫外到近红外光谱辐射计,光谱范围为300 nm~2 000 nm,光谱辐射亮度测量范围为50 μW/cm2·nm·sr~1 000 μW/cm2·nm·sr。阐述了近紫外到近红外光谱辐射计设计原理及关键部件,使用基于钨带灯的直接定标法实现了光谱辐射计光谱辐射亮度绝对定标,测量了标准积分球光源的光谱辐射亮度,测量值与积分球光源标准值偏差优于0.5%。
Abstract
In order to realize high precision measurement of spectral radiance, a kind of small field near ultraviolet(UV) to near infrared(IR) spectroradiometer was developed with a wavelength range of 300 nm~ 2 000 nm and a spectral radiance measurement range of 50 μW/cm2·nm·sr~1 000 μW/cm2·nm·sr. First the design principle and critical components of the near-UV to near-IR spectroradiometer were described. Then the direct method based on tungsten strip lamp was used to realize the spectral radiance absolute calibration of this spectroradiometer. Finally, the spectral radiance of the standard integrated sphere source was measured. The results show that the deviation between measured and standard values is less than 0.5%.
参考文献

[1] 周洪军, 霍同林, 钟鹏飞, 等. 光谱辐射标准和计量实验站的建设和研究[J]. 中国科学技术大学学报, 2007, 37(4): 402-406.

    ZHOU Hongjun, HUO Tonglin, ZHONG Pengfei, et al. Development and research of spectral radiation standard and metrology beamline[J]. Journal of University of Science and Technology of China, 2007, 37(4): 402-406.

[2] 胡铁力, 申越, 郭羽, 等. 低噪声红外辐射计设计[J]. 应用光学, 2013, 34(4): 663-666.

    HU Tieli, SHEN Yue, GUO Yu, et al. Low noise infrared radiometer[J]. Journal of Applied Optics, 2013, 34(4): 663-666.

[3] 范纪红. 光谱辐射参数测试技术的研究[D]. 西安: 陕西师范大学, 2003.

    FAN Jihong. The research of measurement technique about spectral radiation parameters[D]. Xian: Shaanxi Normal University, 2003.

[4] 张恩阳, 闫晓剑, 夏维高. 基于可调法珀腔的微型智能近红外光谱仪研究[J]. 自动化与仪器仪表, 2017(10): 207-209.

    ZHANG Enyang, YAN Xiaojian, XIA Weigao. Research in smart micro-near infrared spectrometer based on ajustable fabry-perot cavity[J]. Automation & Instrumentation, 2017(10): 207-209.

[5] 王学新, 杨鸿儒, 俞兵, 等. 红外目标等立体角标定和测量方法研究[J]. 应用光学, 2018, 39(4): 518-521.

    WANG Xuexin, YANG Hongru, YU Bing, et al. Calibration and measurement method for IR target under the same solid angle[J]. Journal of Applied Optics, 2018, 39(4): 518-521.

[6] 刘加庆, 韩顺利, 刘磊, 等. 一种基于渐变滤光片的2~14μm常温红外光谱辐射测试技术[J]. 红外, 2017, 38(7): 22-30.

    LIU Jiaqing, HAN Shunli, LIU Lei, et al. An 2~14 μm ambient infrared spectroradiometric measurement technology based on linear variable filter[J]. Infrared, 2017, 38(7): 22-30.

[7] 王学新, 杨照金, 岳文龙, 等. 辐射计校准技术的研究[J]. 应用光学, 2004, 25(5): 18-21.

    WANG Xuexin, YANG Zhaojin, YUE Wenlong, et al. The research of calibration of radiometers[J]. Journal of Applied Optics, 2004, 25(5): 18-21.

[8] 郑克哲. 光学计量[M]. 北京: 原子能出版社, 2002.

    ZHENG Kezhe. Optical metrology[M]. Beijing:Atomic Energy Press, 2002.

[9] 贾辉, 李福田. 影响光谱辐亮度标定因素的分析[J]. 光学学报, 2004, 24(3): 383-387.

    JIA Hui, LI Futian. Analysis of factors affecting spectral radiance calibration by applying a diffuse plane[J]. Acta Optica Sinica, 2004, 24(3): 383-387.

[10] 杨小虎, 王淑荣, 曲艺, 等. 基于绝对辐亮度计定标的积分球紫外光谱辐射特性研究[J]. 中国激光, 2011, 38(10): 1008007.

    YANG Xiaohu, WANG Shurong, QU Yi, et al. Research of UV spectral radiance characteristics of the integrating sphere based on calibration by the absolute radiance radiometer[J]. Chinese Journal of Lasers, 2011, 38(10): 1008007.

[11] PATRICK H J , ZAROBILA C J, GERMER T A, et al. The nist robotic optical scatter instrument (ROSI) and its application to BRDF measurements of diffuse reflectance standards for remote sensing[C]. SPIE, 2013,8866(2):1-15.

[12] 邢进, 王淑荣, 李福田. 空间紫外遥感光谱辐射计光谱辐亮度定标三种方法的比较[J]. 中国激光, 2006, 33(4): 509-515.

    XING Jin, WANG Shurong, LI Futian. Comparison of spectral radiance calibrations of spectroradiometer for ultraviolet space remote sensing using three calibration techniques[J]. Chinese Journal of Lasers, 2006, 33(4): 509-515.

[13] SPERFELD P, STOCK K D, RAATZ K H, et al. Characterization and use of deuterium lamps as transfer standards of spectral irradiance[J]. Metrologia, 2003, 40(1): S111-S114.

[14] 范纪红, 赵生禄, 占春连, 等. 成像光谱仪绝对辐射定标技术研究[J]. 应用光学, 2013, 34(4): 629-632.

    FAN Jihong, ZHAO Shenglu, ZHAN Chunlian, et al. Absolute radiometric calibration technique of imaging spectrometer[J]. Journal of Applied Optics, 2013, 34(4): 629-632.

[15] 项金蓉. 景物光谱辐射模拟源研究[D]. 中国科学院研究生院,2014.

    XIANG Jinrong. Research on scenery spectral radiation simulator[D]. Beijing:Chinese Academy of Sciences,2014.

袁林光, 范纪红, 周晓光, 占春连, 卢飞, 李燕, 张博妮, 陈娟, 尤越. 近紫外到近红外光谱辐射计及定标方法研究[J]. 应用光学, 2019, 40(3): 478. YUAN Linguang, FAN Jihong, ZHOU Xiaoguang, ZHAN Chunlian, LU Fei, LI Yan, ZHANG Boni, CHEN Juan, YOU Yue. Research on near-UV to near-IR spectroradiometer and its calibration method[J]. Journal of Applied Optics, 2019, 40(3): 478.

本文已被 4 篇论文引用