强激光与粒子束, 2012, 24 (9): 2230, 网络出版: 2012-09-12   

成像板伽马图像测量研究

Imaging plates applied to gamma-ray image measurement
作者单位
中国工程物理研究院 核物理与化学研究所, 四川 绵阳 621900
摘要
研究了成像板辐射图像测量的基本原理及其物理机制,并将其应用于伽马图像测量。建立了MS,SR和TR三种类型成像板数值模拟模型,分别使用MCNPX程序和基于GEANT4开发的NPE程序计算了三种成像板对不同能量伽马射线的能量沉积,计算结果表明:SR和MS成像板比TR成像板能量沉积在低能部分大3~5倍,高能部分大7~9倍。实验测量了MS成像板γ灵敏度随铜膜厚度的变化关系,测量结果与理论计算结果有较好的一致性。理论与实验结果表明:成像板伽马图像测量的空间分辨力优于50 μm。
Abstract
Basic principles and physical mechanism of image measurement by using imaging plates(IPs) are studied in this paper, and fundamental problems about applications of the IPs in gamma-ray image measurement are also analyzed. Monte Carlo models of imaging plates in common use are built, involving three types of IPs, i.e. MS, SR, TR. And the energy deposition of gamma rays of different energies in the IPs are calculated through the MCNPX code and the NPE code based on a Geant4 platform. As a result, the energy deposition of SR-type and MS-type IPs is 3-5 times larger than that of TR-type IPs in the low energy part, and 7-9 times larger in the high energy part. The theoretical and experimental results of the relation between gamma-ray sensitivity of MS-type IPs and copper foil thickness agree well with each other. The spatial resolution of IPs in gamma-ray imaging is better than 50 μm.
参考文献

[1] Villers D, Paternostre L, Dosière M, et al. The use of an imaging plate as a detecting system in X-ray diffraction of polymers[J].Nucl Instrum Meth B,1995,97:265-268.

[2] 骆志平,Suzuki C,Kosako T,等.成像板对X射线的能量响应[J].原子能科学技术,2009,43(7):644-647.(Luo Zhiping, Suzuki C, Kosako T, et al. Energy response of imaging plate for X-ray. Atomic Energy Science and Technology, 2009,43(7):644-647)

[3] Ohuchi H, Kondo Y. Deeply trapped electrons in imaging plates and their utilization for extending the dynamic range[J].Nucl Instrum Meth A,2010,621(1/3):468-472.

[4] Masaki Y, Takashi K, Tomoya U, et al. Evaluation of high spatial resolution imaging plate[J].Nucl Instrum Meth A,1998,416(2/3):314-318.

[5] Gales S G, Bently C D. Image plates as X-ray detectors in plasma physics experiments[J].Rev Sci Instrum A,2004,75(10):4001-4003.

[6] Amemiya Y. X-ray storage-phosphor imaging-plate detectors: High-sensitivity X-ray area detector[J]. Methods in Enzymology,1997,276:233-243.

[7] Pugliesi R, Lehmann E. Neutron-induced electron radiography using an imaging plate[J].Applied Radiation and Isotopes,2005,62:457-460.

[8] Masao D, Takano J, Inoue M, et al. Applications of imaging plates as position sensitive detectors of positrons[J].Nucl Instrum Meth A,1997,394(1/2):146-150.

[9] Nohtomi A, Terunuma T, Kohno R, et al. Response characteristics of an imaging plate to clinical proton beams[J].Nucl Instrum Meth A,1999,424(2/3):569-574.

[10] Kenji T, Seiji T, Junji M, et al. Imaging performance of imaging plate neutron detectors[J].Nucl Instrum Meth A,1996,377(1):119-122.

[11] Toshiya S, Mamoru B, Keiichiro S, et al. Fast-neutron profiling with an imaging plate[J].Nucl Instrum Meth A,2001,458(3):720-728.

[12] Zawisky M, Hameed F, Dyrnjaja E, et al. Digitized neutron imaging with high spatial resolution at a low power research reactor: I. Analysis of detector performance[J].Nucl Instrum Meth A,2008,587(2/3):342-349.

[13] Chankow N, Punnachaiya S, Wonglee S. Neutron radiography using neutron imaging plate[J].Applied Radiation and Isotopes,2010,68(4/5):662-664.

[14] Sakoda A, Ishimori Y, Hanamoto K, et al. Applicability and performance of an imaging plate at subzero temperatures[J].Applied Radiation and Isotopes,2010,68:2013-2015.

付澜, 张建华, 祁建敏, 章法强, 谢红卫, 陈进川. 成像板伽马图像测量研究[J]. 强激光与粒子束, 2012, 24(9): 2230. Fu Lan, Zhang Jianhua, Qi Jianmin, Zhang Faqiang, Xie Hongwei, Chen Jinchuan. Imaging plates applied to gamma-ray image measurement[J]. High Power Laser and Particle Beams, 2012, 24(9): 2230.

本文已被 3 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!