应用光学, 2015, 36 (3): 408, 网络出版: 2015-09-08   

低温辐射计的结构优化方案设计

Optimization scheme design on construction of cryogenic radiometer
作者单位
1 中国科学院 通用光学定标与表征技术重点实验室, 安徽 合肥 230031
2 中国电子科技集团公司第四十一研究所, 山东 青岛 266555
摘要
传统低温辐射计的布儒斯特窗口会对传递探测器的绝对光谱响应率定标产生较大的影响.文中阐述了低温辐射计对传递探测器的传统定标过程;分析定标过程中因布儒斯特窗口及窗口复现导致定标精度降低的原因;提出一种新的定标结构.该结构中低温辐射计和传递探测器安置于同心弧形轨道上,二者通过精密电机的控制可以分时切入定标光路,实现传递探测器相对于低温辐射计的绝对定标.新的定标结构能够完全消除布儒斯特窗口在传递探测器定标过程中的影响,降低测量激光功率约50%的不确定度,可以进一步降低辐射定标溯源基准的不确定度,提高定标的精度.
Abstract
The Brewster window of traditional cryogenic radiometer has a great influence on the accuracy of absolute spectral responsivity calibration of standard transfer traps.In this paper,the traditional calibration for standard transfer trap was introduced.The reasons of precision degradation due to Brewster window and its state repeatability were analyzed,based on which a novel calibration scheme was proposed.The cryogenic radiometer and standard transfer trap were assembled on a concentric orbit and manipulated by an accurately electromechanical driving mechanism.By alternately switching the two instruments into calibration optical path,the absolute calibration for a standard transfer trap against cryogenic radiometer could be carried out.The new calibration system could completely eliminate the influences caused by Brewster window,would reduce about 50% uncertainty of laser power measurement and further reduce the uncertainty during the calibration process,which is of great significance in improving the accuracy of radiometric calibration.
参考文献

[1] 徐秋云.光谱辐亮度和辐照度响应度系统级定标方法研究[D].合肥:中国科学院安徽精密机械研究所,2010.

    Xu Qiuyun.Research of system-level method forspectral radiance and irradiance responsivity calibrations[D].Hefei:Anhui Institute of Optics and Fine Mechanics(AIOFM)and Chinese Academy of Sciences(CAS),2010.

[2] 李双.基于低温辐射计的宽波段连续光谱辐射定标技术研究[D].合肥:中国科学院安徽光学精密机械研究所,2006.

    Li Shuang.The continuous spectral radiometric calibration at a wide spectrum based on cryogenic radiometer[D].Anhui Institute of Optics and Fine Mechanics(AIOFM)and Chinese Academy of Sciences(CAS),2006.

[3] 范纪红,杨照金,占春连,等.复现坎德拉的测试方法研究[J].应用光学,2002,23(6):36-38.

    Fan Jihong,Yang Zhaojin,Zhan Chunlian,et al.Research on measuring method for candela reappearance[J].Journal of Appled Optics,2002,23(6):36-38.

[4] Hoyt C,Foukal P.Cryogenic radiometers and their application to the metrology[J].Metrologia,1991,28(3):163-167.

[5] Gentile T,Hpuston J,Hardis J,et al.The NIST high accuracy cryogenic radiomter[J],Applied Optics,1996,35(7):1056-1067.

[6] Houston J M,Rice J P.NIST reference cryogenic radiometer designed for versatile performance[J].Metrologia,2006,43(S):31-35.

[7] Carter A C,Lorentz S R,Jung T M,et al.Datla,ACR II:Improved absolute cryogenic radiometer for low background infrared calibrations[J].Applied Optics.2005,44(6):871-875.

[8] Goebel R,Pello R,Ko¨hler R,et al.Comparison of the BIPM cryogenic radiometer with a mechanically cooled cryogenic radiometer from the NPL[J].Metrologia,1996,33:177-179.

[9] Richter M,Kroth U,Gottwald A,et al.Metrology of pulsed radiation for 157nm lithography[J].Applied Optics,2002,41(34):7167-7172.

[10] Kück S,Hofer H,Antonio M,et al.Cryogenic radiometer-based high accurate measurement of Ge and InGaAs trap detector responsivity[C].USA:OSA/CLEO,2006.

[11] 郑小兵,吴浩宇,章骏平,等.不确定度优于0.035%的绝对光谱响应率标准探测器[J].光学学报,2001,21(6):749-752.

    Zheng Xiaobing,Wu Haoyu,Zhang Junping,et al.Absolute spectral responsivity standard detectors with uncertainty less than 0.035%[J].Acta Optica Sinica.2001,21(6):749-752.

[12] 李双,王骥,章骏平,等.可见近红外(488nm~944nm)基于低温辐射计的高精度光辐射绝对定标研究[J].光学学报,2005,25(5):609-613.

    Li Shuang,Wang Ji,Zhang Junping,et al.Highly accurate calibration of optical radiation based on cryogenic radiometer at visible and near infrared spectrum(488nm~944nm)[J].Acta Optica Sinica,2005,25(5):609-613.

[13] Fox N P,Martin J E.Comparison of two cryogenic radiometers by determining the absolute spectral responsivity of silicon photodiodes with an uncertainty of 0.02%[J].Applied Optics ,1990,29(31):4686-4693.

[14] 李健军,郑小兵,卢云君,等.硅陷阱探测器在350nm~1 064nm波段的绝对光谱响应度定标[J].物理学报,2009,58(9):6273-6278.

    Li Jianjun,Zheng Xiaobing,Lu Yunjun,et al.Accurate calibration of the spectral responsivity of silicon trap detectors between 350nm and 1 064nm[J].Chin.Phys.Soc.,2009,58(9):6273-6278.

[15] Atkinson E G.,Butler D J,Calibration of an InGaAs photodiode at 1 300nm with a cryogenic radiometer and a diode laser[J].Metrologia,1998(35):241-245.

[16] Corredera P,Campos J,Hernanz M L,et al.Calibration of near-infrared transfer standards at optical-fibre communication wavelengths by direct comparion with a cryogenic radiometer[J].Applied Optics,1998(35):273-277.

[17] Taylor B N,Kuyatt C E.Guidelines for evaluating and expressing the uncertainty of NIST measurement results[P/OL].2009-07-06[2010-10-05].NIST Technical Note 1297,http:∥physics.nist.gov/ pubs/ guidelines.

庞伟伟, 郑小兵, 李健军, 吴浩宇, 史学舜. 低温辐射计的结构优化方案设计[J]. 应用光学, 2015, 36(3): 408. Pang Weiwei, Zheng Xiaobing, Li Jianjun, Wu Haoyu, Shi Xueshun. Optimization scheme design on construction of cryogenic radiometer[J]. Journal of Applied Optics, 2015, 36(3): 408.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!