光学学报, 2016, 36 (10): 1022001, 网络出版: 2016-10-12  

近红外宽波段下的Czerny-Turner结构复合方法消像散设计

Anastigmatic Design with Czerny-Turner Structure Composite Method on Near-Infrared Broadband
作者单位
1 长春理工大学光电工程学院, 吉林 长春 130022
2 长春理工大学空间光电技术研究所, 吉林 长春 130022
摘要
针对Czerny-Turner结构光谱仪宽波段像散很难同时校正的不足,提出了一种复合方法校正像散,即在一阶消像散方法校正宽波段像散的能力达到极限时,加入柱镜,利用柱镜的相反像散变化趋势,进一步补偿光学系统的剩余像散,推导出复合方法的边缘波段像散补偿公式。应用复合方法,设计了1个近红外900~1700 nm的消像散Czerny-Turner结构。Zemax的仿真结果表明,全波段全视场均方根(RMS)值均小于14 μm,调制传递函数(MTF)达到0.7以上,全波段的光谱分辨率为1.5 nm。保证了在高光谱分辨率的情况下,实现了近红外800 nm宽波段像散的同时校正,避免了能量的横向扩散。该设计方法同样适用其他波段的结构设计,对宽波段消像散型光学系统的设计具有指导意义。
Abstract
A composite method is proposed to correct the astigmatism, which aims at the problem that the astigmatism of the wide band of the Czerny-Turner structure spectrometer is hard to correct at the same time. With the new approach, when the ability of the first-order astigmatism correction reaches the limitation on wide broadband wavelength, the astigmatism opposite trend of added cylindrical lens is used to further compensate the residual astigmatism in the optical system. The astigmatism compensation formula on marginal waveband of with composite method is deduced. With the composite method, an anastigmatic Czerny-Turner structure on near-infrared waveband ranging from 900 nm to 1700 nm is designed. The simulation results of Zemax show that the root-mean-square (RMS) value of full-band and full fields of view is less than 14 μm, modulation transmission function (MTF) is higher than 0.7, and the spectral resolution is 1.5 nm at all wavelengths. The composite method achieves simultaneous correction of astigmatism at near-infrared 800 nm wide waveband, which can avoid energy transverse spread. The method can be applied in structure design at other wavelengths, and it is meaningful for designing broadband anastigmatic optical systems.
参考文献

[1] 孙慈, 姚雪峰, 崔继承, 等. 基于短波红外成像光谱仪的矿石光谱测量[J]. 光学学报, 2016, 36(2): 0230001.

    Sun Ci, Yao Xuefeng, Cui Jicheng, et al. Mineral spectrum measurement based on shortwave infrared imaging spectrometer[J]. Acta Optica Sinica, 2016, 36(2): 0230001.

[2] 刘兵, 刘英, 李灿, 等. 轻小型可见/近红外实时成像光谱仪的光学系统设计[J]. 光学学报, 2015, 35(6): 0630002.

    Liu Bing, Liu Ying, Li Can, et al. Design of optical system for miniature visible/near-infrared real-time imaging spectrometer[J]. Acta Optica Sinica, 2015, 35(6): 0630002.

[3] 赵敏杰, 司福祺, 陆亦怀, 等. 星载大气痕量气体差分吸收光谱仪光谱图像校正方法研究[J]. 光学学报, 2015, 35(12): 1230002.

    Zhao Minjie, Si Fuqi, Lu Yihuai, et al. Spectral image correction method for space-born differential optical absorption spectrometer[J]. Acta Optica Sinica, 2015, 35(12): 1230002.

[4] 陈洪耀, 张黎明, 李鑫, 等. 高光谱遥感器飞行中基于大气特征谱线的光谱定标技术[J]. 光学学报, 2013, 33(5): 0528003.

    Chen Hongyao, Zhang Liming, Li Xin, et al. Hyperspectral sensor in flight spectral calibration based on characteristic spectra of atmosphere[J]. Acta Optica Sinica, 2013, 33(5): 0528003.

[5] Bates B, McDowell M, Newton A C. Correction of astigmatism in a Czerny-Turner spectrograph using a plane grating in divergent illumination[J]. Journal of Physics E: Scientific Instruments, 1970, 3(3): 206-210.

[6] Masayuki F, Takehiko T, Kei-ichi K. Highly efficient and aberration-corrected spectrometer for advanced Raman spectroscope[J]. Applied Optics, 2002, 41(22): 4655-4665.

[7] Xu L, Chen K X, He Q S, et al. Design of freeform mirrors in Czerny-Turner spectrometers to suppress astigmatism[J]. Applied Optics, 2009, 48(15): 2871-2879.

[8] Austin D R, Witting T, Walmsley I A. Broadband astigmatism-free Czerny-Turner imaging spectrometer using spherical mirrors[J]. Applied Optics, 2009, 48(19): 3846-3853.

[9] 薛庆生, 陈伟. 改进的宽波段车尔尼-特纳光谱成像系统设计[J]. 光学 精密工程, 2012, 20(2): 233-240.

    Xue Qingsheng, Chen Wei. Design of modified Czerny-Turner spectral imaging system with wide spectral region[J]. Optics and precision Engineering, 2012,20(2): 233-240.

[10] Lee K S, Thompson Kevin P, Rolland J P. Broadband astigmatism-corrected Czerny-Turner spectrometer[J]. Optics Experess, 2010, 18(22): 23378-23384.

[11] Xue Q S. Astigmastism-corrected Czerny-Turner imaging spectrometer for broadband spectral simultaneity[J]. Applied Optics, 2011, 50(10): 1338-1344.

[12] Qu B X, Guo X, Yu F H. Optimization of freeform cylindrical lens in Czerny-Turner spectrometer system to correct the astigmastism by using user defined error function[C]. SPIE, 2012, 8416: 84161I.

[13] 刘健鹏, 唐义, 黄刚, 等. 改进型Czerny-Turner成像光谱仪光学系统设计方法[J]. 光学学报, 2012, 32(3): 0322007.

    Liu Jianpeng, Tang Yi, Huang Gang, et al. Design method of optical system of improved Czerny-Turner imaging spectrometer[J]. Acta Optica Sinica, 2012, 32(3): 0322007.

李欣航, 董科研, 安岩. 近红外宽波段下的Czerny-Turner结构复合方法消像散设计[J]. 光学学报, 2016, 36(10): 1022001. Li Xinhang, Dong Keyan, An Yan. Anastigmatic Design with Czerny-Turner Structure Composite Method on Near-Infrared Broadband[J]. Acta Optica Sinica, 2016, 36(10): 1022001.