Novel calibration optical path of cryogenic radiometer Download: 1126次
1 Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei 230031, China
2 The 41st Institute of China Electronics Technology Group Corporation, Qindao 266555, China
Figures & Tables
Fig. 1. Laser power measurement based on cryogenic radiometer.
下载图片 查看原文
Fig. 2. Transfer detector on calibration of cryogenic radiometer.
下载图片 查看原文
Fig. 3. Novel structure on the calibration of cryogenic radiometer.
下载图片 查看原文
Fig. 4. Transfer detector installed in detector cavity.
下载图片 查看原文
Fig. 5. Optical path on the calibration of cryogenic radiometer.
下载图片 查看原文
Table1. Uncertainty on Measuring the Laser Power of Cryogenic Radiometera
| Wavelength (nm) |
---|
Sources | 488 | 514 | 633 | 676 | 786 | 830 | 944 |
---|
Electrical power | 0.31 | 0.362 | 0.389 | 0.296 | 0.260 | 0.174 | 0.140 | Window transmittance | 1.525 | 0.609 | 1.529 | 0.437 | 0.453 | 0.521 | 0.520 | Receiver absorptance | | | | 0.1 | | | | Heating nonequivalence | | | | 0.087 | | | | Total uncertainty | 1.541 | 0.652 | 1.548 | 0.488 | 0.539 | 0.565 | 0.555 | Percent | 98.96 | 93.40 | 98.77 | 89.55 | 84.04 | 92.21 | 93.69 |
|
查看原文
Table2. Technique Parameter of Cryogenic Radiometer at the Anhui Institute of Optics and Fine Mechanics
Technical Specifications | |
---|
Instrument model | Cryorad II | Spectrum | 0.25–50 μm | Receiver response | | Dynamic range | Max to 0.25 mW | Absolute accuracy | | Receiver absorptance | (632.8 nm) | Refrigeration mode | Liquid nitrogen | | Liquid helium | Calibration structure | Y-type | Demount Brewster window | No |
|
查看原文
Table3. Comparison on Responsivity of Transfer Detector in Novel Calibration Structure and Traditional Calibration Structure
| Absolute Spectral Responsivity (A/W) |
---|
Detector | Traditionala (2008) | Novel (2014) | Trap-B | 0.5069 | 0.5045 | K | 0.4% |
|
查看原文
Table4. Compare of Combine Uncertainty of Transfer Detector between Traditional and Novela
Source | Traditional Calibration Structure (2008)b | Novel Calibration Optical Path (2014) |
---|
Electrical power | 0.389 | 0.671 | Window transmittance | 1.529 | — | Receiver absorptance | 0.1 | 0.1 | Heating nonequivalence | 0.087 | 0.05195 | Laser power total uncertainty | 1.5832 | 0.6804 | Linearity | 0.434 | Spatial uniformity | 1.99 | Polarization sensitivity | 0.492 | Stability | 0.687 | Output V | 0.025 | 0.0244 | Total uncertainty () | 2.715 | 2.321 |
|
查看原文
Weiwei Pang, Xiaobing Zheng, Jianjun Li, Xueshun Shi, Haoyu Wu, Maopeng Xia, Dongyang Gao, Jianmin Shi, Tao Qi, Qing Kang. Novel calibration optical path of cryogenic radiometer[J]. Chinese Optics Letters, 2015, 13(5): 051201.