Chinese Optics Letters, 2015, 13 (5): 051201, Published Online: Sep. 20, 2018   

Novel calibration optical path of cryogenic radiometer Download: 1126次

Author Affiliations
1 Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei 230031, China
2 The 41st Institute of China Electronics Technology Group Corporation, Qindao 266555, China
Figures & Tables

Fig. 1. Laser power measurement based on cryogenic radiometer.

下载图片 查看原文

Fig. 2. Transfer detector on calibration of cryogenic radiometer.

下载图片 查看原文

Fig. 3. Novel structure on the calibration of cryogenic radiometer.

下载图片 查看原文

Fig. 4. Transfer detector installed in detector cavity.

下载图片 查看原文

Fig. 5. Optical path on the calibration of cryogenic radiometer.

下载图片 查看原文

Table1. Uncertainty on Measuring the Laser Power of Cryogenic Radiometera

Wavelength (nm)
Sources488514633676786830944
Electrical power0.310.3620.3890.2960.2600.1740.140
Window transmittance1.5250.6091.5290.4370.4530.5210.520
Receiver absorptance0.1
Heating nonequivalence0.087
Total uncertainty1.5410.6521.5480.4880.5390.5650.555
Percent98.9693.4098.7789.5584.0492.2193.69

查看原文

Table2. Technique Parameter of Cryogenic Radiometer at the Anhui Institute of Optics and Fine Mechanics

Technical Specifications 
Instrument modelCryorad II
Spectrum0.25–50 μm
Receiver response2KmW1
Dynamic rangeMax to 0.25 mW
Absolute accuracy±0.005%
Receiver absorptance>0.99998 (632.8 nm)
Refrigeration modeLiquid nitrogen
Liquid helium
Calibration structureY-type
Demount Brewster windowNo

查看原文

Table3. Comparison on Responsivity of Transfer Detector in Novel Calibration Structure and Traditional Calibration Structure

 Absolute Spectral Responsivity (A/W)
DetectorTraditionala (2008)Novel (2014)
Trap-B0.50690.5045
K0.4%

查看原文

Table4. Compare of Combine Uncertainty of Transfer Detector between Traditional and Novela

SourceTraditional Calibration Structure (2008)bNovel Calibration Optical Path (2014)
Electrical power0.3890.671
Window transmittance1.529
Receiver absorptance0.10.1
Heating nonequivalence0.0870.05195
Laser power total uncertainty1.58320.6804
Linearity0.434
Spatial uniformity1.99
Polarization sensitivity0.492
Stability0.687
Output V0.0250.0244
Total uncertainty (k=1)2.7152.321

查看原文