A convex aspheric surface using a computer-generated hologram (CGH) test plate fabricated with novel techniques and equipment is tested. However, the measurement result is not verified via comparison with other methods. To verify the accuracy of the measurement, a perfect sphere surface is measured by the following. The measurement result is quantified into four parts: the figure error from the tested spherical surface; the figure error from the reference spherical surface; the error from the hologram; and the adjustment error from misalignment. The measurement result, removed from the later three errors, shows agreement to 4-nm RMS with the test by Zygo interfermeter of the same surface. Analysis of the CGH test showed the overall accuracy of the 4-nm RMS, with 3.9 nm from the test plate figure, 0.5 nm from the hologram, and 0.74 nm from other sources, such as random vibration, various second order effects, and so on. Thus, the measurement accuracy using the proposed CGH could be very high. CGH can therefore be used to measure aspheric surfaces accurately.
References

[1] A. Offner and D. Malacara, Optical Shop Testing (Wiley, New York, 2007).

[2] M. V. Mantravadi, V. Kumar, and R. J. von Handorf, Proc. SPIE 1332, 107 (1990).

[3] J. H. Burge, Proc. SPIE 2576, 258 (1995).

[4] J. H. Burge and D. S. Anderson, Proc. SPIE 2199, 181 (1994).

[5] Y. Xie, Z. Lu, F. Li, J. Zhao, and Z. Weng, Opt. Express 10, 1043 (2002).

[6] J. Ma, Z. Gao, R. Zhu, Y. He, L. Chen, J. Li, E. Pun, W. Wong, L. Huang, and C. Xie, Chin. Opt. Lett. 7, 70 (2009).

[7] Y. Xie, Z. Lu, and F. Li, Opt. Express 11, 992 (2003).

[8] Y. Xie, Z. Lu, and F. Li, Opt. Express 11, 975 (2003).

[9] Y. Xie, Z. Lu, and F. Li, Opt. Express 12, 1810 (2004).

[10] H. Liu, Z. Lu, and F. Li, Opt. Laser Technol. 37, 642 (2005).

[11] Y. Chang and J. H. Burge, Proc. SPIE 3782, 358 (1999).

[12] H. Liu, Z. Lu, F. Li, Y. Xie, S. Kan, and S. Wang, Opt. Express 12, 3251 (2004).

[13] H. Liu, Z. Lu, F. Li, and Y. Xie, Opt. Commun. 242, 339 (2004).

[14] Z. Lu and Z. Weng, Proc. SPIE 4093, 176 (2000).

Hua Liu. Measurement accuracy verification of aspheric surface test with computer-generated hologram[J]. Chinese Optics Letters, 2012, 10(7): 071201.

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!