Chinese Optics Letters, 2015, 13 (9): 093401, Published Online: Sep. 14, 2018
Application of polycapillary x ray lens to eliminate both the effect of x ray source size and scatter of the sample in laboratory tomography Download: 893次
Figures & Tables
Fig. 1. Scheme of the x ray absorption imaging device based on a polycapillary x ray lens.
Fig. 2. (a) Picture of the PPXRL. (b) Picture of the PCXRL. (c) Cross section of the outside of the PPXRL. The channel size is about 7 μm.
Fig. 4. X ray intensity distribution of the output beam of the PPXRL at a distance of 30 cm from the CCD detector.
Fig. 5. Distribution of the gain of the PPXRL at a distance of 30 cm from the CCD detector.
Fig. 7. Image of the mesh at a distance of 20 cm from the PPXRL taken by the tomography device (a) with and (b) without the PCXRL.
Fig. 9. (a) Ratio of intensities in Fig. 8 . (b) Image of Fig. 7 after flat-field correction.
Fig. 10. Image of the mesh in different places. (a) L 1 = 20 cm and L 2 = 2 cm . (b) L 1 = 20 cm and L 3 = 2 cm .
Fig. 11. Contrast of the images as a function of the distance between the detector and the outside of PCXRL (the red line), and between the sample and the inside of PCXRL (the black line).
Table1. The Main Geometric Parameters of the PPXRL and PCXRL
|
Xuepeng Sun, Zhiguo Liu, Tianxi Sun, Longtao Yi, Weiyuan Sun, Fangzuo Li, Bowen Jiang, Yongzhong Ma, Xunliang Ding. Application of polycapillary x ray lens to eliminate both the effect of x ray source size and scatter of the sample in laboratory tomography[J]. Chinese Optics Letters, 2015, 13(9): 093401.