光电工程, 2016, 43 (6): 68, 网络出版: 2016-07-26   

双波长视网膜血氧测量系统

Dual-wavelength Retinal Oximeter
先永利 1,2,3,4,*戴云 1,3高椿明 2杜睿 1,3
作者单位
1 中国科学院自适应光学重点实验室,成都 610209
2 电子科技大学光电信息学院,成都 610054
3 中国科学院光电技术研究所,成都 610209
4 中国科学院大学,北京 100049
摘要
为了测量视网膜血氧饱和度,本文设计一种以商业眼底相机为平台的双波长视网膜血氧测量系统。该系统将商业眼底相机输出的视网膜像进行二次成像,同时采集含氧血红蛋白和还原血红蛋白的等吸收波段和非等吸收波段图像。对获取的双波长图像进行图像配准、血管分割、光密度比计算等一系列运算,最终完成视网膜血管血氧饱和度的测量。对健康人眼进行重复性测量,动脉和静脉的血氧饱和度平均值分别是 92.85%和 56.75%,标准差的平均值分别为 2.41%和 3.63%,表明系统的准确性和重复性较好。本系统在获取视网膜结构图像的同时获得视网膜血管功能信息,可以为生命科学相关研究、眼底相关疾病的诊断提供有力工具。
Abstract
In order to measure retinal oxygen saturation, a dual-wavelength retinal oximeter based on commercial fundus camera was designed. The system made the retinal image of the commercial fundus camera second imaging, and simultaneously collected dual-wavelength images that respectively represented the same absorption coefficient wavelength images and the different one of oxygenated hemoglobin and de-oxygenated hemoglobin. The retinal oxygen saturation was finally measured after image registration, vessel segmentation and calculation of optic density ratio of the two images. The means oxygen saturation of artery and venules in health human for repeated measurements were 92.85% and 56.75%, respectively. The mean standard deviation was 2.41% for arterioles and 3.63% for venules.The data demonstrate that the system is reliable and reproducible. The system can obtain retinal structural imaging, which provides a useful tool for research of life sciences and diagnosis of retinal-related diseases.
参考文献

[1] Hardarson S H,Stefansson E. Retinal oxygen saturation is altered in diabetic retinopathy [J]. British Journal of Ophthalmology(S1468-2079),2012,96(4):560-563.

[2] Hammer M,Heller T,Jentsch J,et al. Retinal vessel oxygen saturation under flicker light stimulation in patients with nonproliferative diabetic retinopathy [J]. Investigative Ophthalmology & Visual Science(S1552-5783),2012,53(7): 4063-4068.

[3] Hardarson S H,Gottfredsdottir M S,Halldorsson G H,et al. Glaucoma filtration surgery and retinal oxygen saturation [J]. Investigative Ophthalmology & Visual Science(S1552-5783),2009,50(11):5247-5250.

[4] Evelien Vandewalle,Luis Abegao Pinto,Olof Birna Olafsdottir,et al. Phenylephrine 5% added to Tropicamide 0.5% eye drops does not influence retinal oxygen saturation values or retinal vessel diameter in glaucoma patients [J]. Acta Ophthalmologica(S1755-3768),2013,91(8):733-737.

[5] Olof Birna Olafsdottir,Sveinn Hakon Hardarson,Maria Soffia Gottfredsdottir,et al. Retinal Oximetry in Primary Open-Angle Glaucoma [J]. Investigative Ophthalmology & Visual Science(S1552-5783),2011,52(9):6409-6413.

[6] Sveinn Hakon Hardarson,Einar Stefansson. Oxygen saturation in central retinal vein occlusion [J]. American Journal of Ophthalmology(S0002-9394),2010,150(6):871-875.

[7] Hardarson S H,Elfarsson A,Agnarsson A B,et al. Retinal oximetry in central retinal artery occlusion [J]. Acta Ophthalmologica(S1755-3768),2013,91(2):189-190.

[8] Hardarson S H. retinal oximetry [D]. Reykjavík:University of Iceland,2013:1-7.

[9] Hickham J B,Frayser R,Ross J C. A study of retinal venous blood oxygen saturation in human subjects by photographic means [J]. Circulation(S1524-4539),1963,27(3):375-385.

[10] Harris A,Dinn R B,Kagemann L,et al. A review of methods for human retinal oximetry [J]. Ophthalmic Surgery Lasers & Imaging(S1542-8877),2003,34(2):152-164.

[11] Beach J M,Schwenzer K J,Srinivas S,et al. Oximetry of retinal vessels by dual-wavelength imaging:calibration and influence of pigmentation [J]. Journal of Applied Physiology(S1522-1601),1999,86(2):748-758.

[12] Hardarson S H,Harris A,Karlsson R A,et al. Automatic retinal oximetry [J]. Investigative Ophthalmology & Visual Science(S1552-5783),2006,47(11):5011-5016.

[13] Nakamura D,Matsuoka N,Tatsuguchi K,et al. Oxygen Saturation Imaging of Human Retinal Vessels and Measurement in Eye Disease Patient for Clinical Application [J]. Proceedings of SPIE(S0277-786X),2010,7550(1):75501x-1-75501x-8.

[14] Bahram Khoobehi,James M Beach,Hiroyuki Kawano. Hyperspectral Imaging for Measurement of OxygenSaturation in the Optic Nerve Head [J]. Investigative Ophthalmology & Visual Science(S1552-5783),2004,45(5):1464-1472.

[15] Dietrich Schweitzer,Martin Hammer,Janine Kraft,et al. In Vivo Measurement of the Oxygen Saturation of Retinal Vessels in Healthy Volunteers [J]. IEEE Transactions on Biomedical Engineering(S0018-9294),1999,46(12):1454-1465.

[16] 张二虎,卞正中,俞江 . 基于互信息的视网膜眼底图像配准方法 [J].西安理工大学学报, 2002,18(1):14-17. ZHANG Erhu,BIAN Zhengzhong,YU Jiang. Mutual Information Based Registration for Retinal Fundus Images [J]. Journal of Xi'an University of Technology,2002,18(1):14-17.

[17] 高原. 高效的医学眼底图像定位与配准研究 [D].上海:上海交通大学,2012:35-41. GAO Yuan. A Study on Fast Localization and Registering of Retinal Fundus Images [D]. Shanghai:Shanghai Jiao Tong University,2012:35-41.

[18] 陈丽平. 基于 Hessian矩阵的血管图像增强与水平集分割算法研究 [D].长沙:湖南大学, 2012:8-15. CHEN Liping. Researches on Vessel Image Enhancement Based on Hessian Matrix and Level-Set Segmentation Algorithm [D]. Changsha:Hunan University,2012:8-15.

[19] 游嘉,陈波 . 基于矩阵的多尺度视网膜图像增强方法 [J].计算机应用, 2011,31(6):1560-1562. YOU Jia,CHEN Bo. New approach to retinal image enhancement based on Hessian matrix [J]. Journal of Computer Applications,2011,31(6):1560-1562.

[20] Hammer M,Vilser W,Riemer T,et al. Retinal vessel oximetry-calibration,compensation for vessel diameter and fundus pigmentation,and reproducibility [J]. Journal of Biomedic Optics(S1083-3668),2008,13(5):054015-1.

[21] Geirsdottir A,Palsson O,Hardarson S,et al. Retinal vessel oxygen saturation in healthy individuals [J]. Investigative Ophthalmology & Visual Science(S1552-5783),2012,53(9):5433-5442.

先永利, 戴云, 高椿明, 杜睿. 双波长视网膜血氧测量系统[J]. 光电工程, 2016, 43(6): 68. XIAN Yongli, DAI Yun, GAO Chunming, DU Rui. Dual-wavelength Retinal Oximeter[J]. Opto-Electronic Engineering, 2016, 43(6): 68.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!