光学学报, 2008, 28 (12): 2325, 网络出版: 2008-12-17   

基于参变下转换效应的光辐射定标方法

Optical Radiometric Calibration Based on Parametric Down-Conversion Effect
作者单位
中国科学院安徽光学精密机械研究所遥感研究室, 合肥 安徽 230031
摘要
介绍了一种新型的基于参变下转换效应的光辐射绝对定标方法, 利用固体钛宝石激光器倍频后产生的351 nm的激光光源抽运非线性BBO晶体, 制备了一对简并的参变下转换光子对(波长为702 nm), 并利用晶体的角度调谐特性, 制备了另一对非简并的参变下转换光子对(波长为633 nm和789 nm)。定标了光电倍增管的量子效率, 介绍了实验原理和定标装置, 系统分析了测量过程中引入的不确定度。结果表明, 参变下转换效应光辐射定标方法的相对合成不确定度小于5.5%。参变下转换定标方法与传统的基于标准探测器的辐射定标方法作了一个比对, 表明两种定标方法相对误差小于10%。
Abstract
A new calibration method of detectors by using correlated photons generated in spontaneous parametric down-conversion effect of nonlinear crystal is proposed. A CW ultraviolet (351 nm), diode-pumped, frequency-doubled, solid-state laser is used to pump BBO crystal. The degenerate parametric down-conversion (PDC) photons with wavelengths of 702 nm were produced and by tuning crystal angle, nondegenerate PDC photons with wavelengths of 633nm and 789nm were also produced. The quantum efficiencies of the photomultiplier at the above wavelengths were measured respectively. An absolute calibration system of detector quantum efficiency was performed. And its principle and experimental setup were introduced. The calibration uncertainties of PDC method were analyzed systematically. The results show the total calibration relative uncertainties are less than 5.5%. In the end, a comparison between PDC effect method and conventional method based on standard detectors was performed. The results show that the relative difference of the two methods is less than 10%.
参考文献

[1] . . High-accuracy primary and transfer standards for radiometric calibration[J]. Chinese Science Bulletin, 2000, 45(21): 2009-2013.

[2] 郑小兵, 吴浩宇, 章骏平 等. 不确定度优于0.035%的绝对光谱[J]. 光学学报, 2000, 21(6): 749~752

    Zheng Xiaobing, Wu Haoyu, Zhang Junping et al.. Absolute spectral responsivity standard detectors with uncertainty less than 0.035%[J]. Acta Optica Sinica, 2001, 21(6): 749~752

[3] Feng Yu, Zheng Xiaobing, Li Jianjun et al.. Research in absolute calibration of single photon detectors by means of correlated photons[J]. Chin. Opt. Lett., 2006, 4(6): 315~317

[4] J. Y. Cheung, M. P. Vaughan, J. R. Mountford et al.. Correlated photon metrology of detectors and sources[C]. Proc. SPIE, 2006, 5161: 365~376

[5] . V. Polyakov, A. L. Migdall. High accuracy verification of a correlated photon-based method for determining photon-counting detection efficiency[J]. Opt. Exp., 2007, 15(4): 1390-1407.

[6] . Brida, S. Castelletto, I. P. Degiovanni et al.. Quantum efficiency and dead time of single-photon counting photodiodes: a comparison between two measurement techniques[J]. Metrologia, 2000, 37(5): 625-628.

[7] 冯瑜, 郑小兵, 张伟 等. 基于相关光子的光辐射绝对定标方法研究[J]. 光学学报, 2007, 27(3): 457~460

    Feng Yu, Zheng Xiaobing, Zhang Wei et al.. Optical radiometric calibration based on correlated photons[J]. Acta Optica Sinica, 2007, 27(3): 457~460

[8] . N. Klyshko. Use of two-photon light for absolute calibration of photoelectric detectors[J]. Sov. Quant. Electron., 1980, 10(11): 1112-1116.

[9] B. N. Tayor, C. E. Kuyatt. Guidelines for evaluating and expressing the uncertainty of NIST measurement results[R]. MD 20899-0001, 1994

[10] . V. Polyakov, A. L. Migdall. High accuracy verification of a correlated photon-based method for determining photon-counting detection efficiency[J]. Opt. Exp., 2007, 15(4): 1390-1407.

[11] 费页泰. 误差理论与数据处理[M]. 北京: 机械工业出版社, 2004

    Fei Yetai. Error Theory and Data Processing[M]. Beijing: China Machine Press, 2004

李健军, 郑小兵, 卢云君, 张伟, 邹鹏. 基于参变下转换效应的光辐射定标方法[J]. 光学学报, 2008, 28(12): 2325. Li Jianjun, Zheng Xiaobing, Lu Yunjun, Zhang Wei, Zou Peng. Optical Radiometric Calibration Based on Parametric Down-Conversion Effect[J]. Acta Optica Sinica, 2008, 28(12): 2325.

本文已被 7 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。