强激光与粒子束, 2012, 24 (9): 2235, 网络出版: 2020-12-21  

基于卷积法TV模型的锥束CT散射校正方法

Scatter correction algorithm for cone-beam computed tomography based on convolution and total variation model
作者单位
信息工程大学 电子科学与技术系, 郑州 450002
摘要
为了解决CT图像中散射伪影问题,提出并使用修正的泊松分布函数作为散射函数,然后利用卷积模型生成多个散射基图像,将基图像与原始重建图像的线性组合作为最终重建图像,最后,通过使重建图像的全变分函数最小求取线性组合系数最优解。该算法具有计算复杂度小、能够灵活应用于不同的成像对象的特点。实验结果表明,算法能够有效抑制散射导致的杯状和黑色带状伪影,提高图像对比度。
Abstract
In order to solve the problem of scattering artifacts in the computed tomography images, a modified Poisson distribution function is presented and used as the scatter model. A set of scatter basis images are obtained from scatter estimations with different parameter values in the model. Then the corrected image is obtained by adding the basis images to the original image. The coefficients for the linear combination are determined by minimizing the total variation of the combined images. The method requires low computational effort and allows flexible application to different scanned objects. Experiment results indicate that cupping and dark bands artifacts due to scattering are effectively depressed by the proposed method, and the image contrast is enhanced.
参考文献

[1] Zhu L, Bennett N R, Fahrig R. Scatter correction method for X-ray CT using primary modulation: theory and preliminary results[J]. IEEE Trans on Medical Imaging, 2006, 25(12):1573-1587.

[2] 王珏, 陈教泽, 谭辉, 等. 工业CT光电二极管探测器灵敏度[J]. 强激光与粒子束, 2012, 24(1):210-214.(Wang Jue, Chen Jiaoze, Tan Hui, et al. Sensitivity of photodiode detector for industrial computed tomography. High Power Laser and Particle Beams, 2012, 24(1):210-214)

[3] Ohnesorge B, Flohr T, Klingenbeck-Regn K. Efficient object scatter correction algorithm for third and fourth generation CT scanners[J]. European Radiology, 1999, 9:563-569.

[4] Siewerdsen J H, Moseley D J, Bakhtiar B, et al. The influence of anti-scatter grids on soft-tissue detect ability in cone-beam computed tomography with flat-panel detectors[J]. Medical Physics,2004, 31:3506-3520.

[5] Naimuddin S, Hasegawa B H, Mistretta C A. Scatter-glare correction using a convolution algorithm with variation weighting[J]. Medical Physics, 1987,14:330-334.

[6] Star-Lack J, Sun M, Kaestner A, et al. Efficient scatter correction using a symmetric kernels[C]//Proc of SPIE. 2009:7258.

[7] Leliveld C J, Maas J G, Bom V R, et al. Monte Carlo modeling of coherent scattering: influence of interference[C]//IEEE Nuclear Science Symposium and Medical Imaging Conference Record.1995, 3:3315-3321.

[8] Kak A C, Slaney M. Principles of computerized tomography[M]. United States:IEEE Service Center, 1987.

[9] Esther M, Clemens M, Matthias B,et al. Empirical scatter correction (ESC):A new CT scatter correction method and its application to metal artifact reduction[C]//IEEE Nuclear Science Symposium Conference Record. 2010, 11:2036-2041.

李庆亮, 闫镔, 李磊, 孙红胜, 张峰. 基于卷积法TV模型的锥束CT散射校正方法[J]. 强激光与粒子束, 2012, 24(9): 2235. Li Qingliang, Yan Bin, Li Lei, Sun Hongsheng, Zhang Feng. Scatter correction algorithm for cone-beam computed tomography based on convolution and total variation model[J]. High Power Laser and Particle Beams, 2012, 24(9): 2235.