光学学报, 1994, 14 (7): 736, 网络出版: 2007-08-17
线阵CCD转移效率及响应度非均匀性测试装置研究
An apparatus for measurement of charge transfer efficiency and photo-response non-uniformity of linear CCD
摘要
介绍了准确测量线阵CCD转移效率及响应度非均匀性的实验方法、装置及测试结果.采用小光点注入法,在自动化装置上实现数据采集,用移动寻峰和直线拟合法处理数据.文章还分析了影响测量结果的诸多因素.
Abstract
The accurate measuring result of charge transfer efficiency (CTE) and photoresponse non-uniformity (PRNU) of linear CCD image sensors have been describedin this paper. A automatic measuring system, bssed on the method of focuSed opt sPOt has been set up. This method of data acquisition and calculation is newish. Factors affecting the characteristic Parameters of CCD are analyzed.
参考文献
万光毅, 龚晓斌, 罗镇伟, 苏昌林, 胡章鑫, 景峰. 线阵CCD转移效率及响应度非均匀性测试装置研究[J]. 光学学报, 1994, 14(7): 736. 万光毅, 龚晓斌, 罗镇伟, 苏昌林, 胡章鑫, 景峰. An apparatus for measurement of charge transfer efficiency and photo-response non-uniformity of linear CCD[J]. Acta Optica Sinica, 1994, 14(7): 736.