Chinese Optics Letters, 2014, 12 (12): 121201, Published Online: Dec. 1, 2014  

Research on calibration method in lab of direct solar channels of Sun photometer Download: 597次

Author Affiliations
1 Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei 230031, China
2 Institute of Remote Sensing Applications, Chinese Academy of Science, Beijing 100101, China
Abstract
We develop a new calibration method in lab by measuring the absolute spectral irradiance responsivity of Sun photometer sun channel. The absolute power responsivity of Sun photometer is obtained when a white laser double monochromator system serve as a source, and a standard transfer detector calibrated against cryogenic absolute radiometer is assembled to measure the absolute power of laser beam. The effective area of aperture is measured through laser raster scanning method, and the relative spectral irradiance responsivity of the corresponding channel is obtained by using tungsten–halogen lamps double monochromator system. On the basis of the above results, the top of the atmosphere responsive constants V0 (500, 675, and 870 nm) are obtained by integration with extraterrestrial solar spectral irradiance data. Comparing the calibration results with that of CIMEL, France in November 2011, the relative differences are 4.38%, 2.23%, and 2.45%, respectively. The calibration uncertainty reaches to 2.048×10-2, which shows a remarkable consistency with the Langley plot method. Further, our scheme can overcome the limits of space and atmospheric conditions which are only available at a high-altitude calibration site in particular date. The advantages lie in not only shortening the experiment period but also being of high precision. This new scheme definitely plays an important role in supporting the current and future sun photometry calibration activities which are significant to earth observation.
References

[1] B. Holben, T. Eck, I. Slutsker, D. Tanre, J. Buis, A. Setzer, E. Vermote, J. Reagan, Y. Kaufman, and T. Nakajima, Remote Sens. Environ. 66, 1 (1998).

[2] Z. Li, L. Blarel, T. Podvin, P. Goloub, J. P. Buis, and J. P. Morel, Appl. Opt. 47, 1368 (2008).

[3] G. E. Shaw, J. A. Reagan, and B. M. Herman, J. Appl. Meteorol. 12, 374 (1973).

[4] X. Wang, Y. Qiao, P. Goloub, and Z. Li, Acta Opt. Sin. 28, 87 (2008).

[5] B. Schmid and C. Wehrli, Appl. Opt. 34, 4500 (1995).

[6] M. Campanelli, T. Nakajima, and B. Olivieri, Appl. Opt. 43, 651 (2004).

[7] M. Chen, J. Davis, H. Tang, Z. Gao, and W. Gao, Proc. SPIE 8513, 851305 (2012).

[8] R. Tao, H. Che, Q. Chen, Y. Wang, J. Sun, X. Zhang, S. Lu, J. Guo, H. Wang, and X. Zhang, Particuology 13, 88 (2014).

[9] Q. Xu, X. Zheng, W. Zhang, and X. Wang, Acta Opt. Sin. 30, 1337 (2010).

[10] W. Zhai, J. Li, X. Zheng, Z. Li, B. Meng, W. Zhang, and X. Wang, Acta Opt. Sin. 32, 131 (2012).

[11] W. Xu, J. Li, and X. Zheng, Spectrosc. Spectr. Anal. 33, 255 (2013).

[12] V. Ahtee, S. W. Brown, T. C. Larason, K. R. Lykke, E. Ikonen, and M. Noorma, Appl. Opt. 46, 4228 (2007).

[13] E. Ikonen, P. Toivanen, and A. Lassila, Metrologia 35, 369 (1998).

[14] J. Sheng, L. Zhang, X. Zheng, S. Li, Infrared Laser Eng. 37, 534 (2008).

[15] X. Zheng, H. Wu, J. Zhang, Y. Liu, W. Zhou, L. Wang, and Y. Qiao, Chin. Sci. Bull. 45, 2009 (2000).

[16] J. Li, X. Zheng, Y. Li, W. Zhang, and P. Zhou, Acta Phys. Sin. 58, 6273 (2009).

[17] J. B. Fowler, R. Durvasula, and A. Parr, Metrologia 35, 497 (1998).

[18] J. T. Woodward, A. W. Smith, C. A. Jenkins, C. Lin, S. W. Brown, and K. R. Lykke, Metrologia 46, S277 (2009).

[19] A. Berjón, B. Torres, C. Toledano, T. Podvin, L. Blarel, N. Prats, P. Goloub, and V. Cachorro, J. Aerosol Sci. 59, 1 (2013).

[20] G. Thuillier, L. Floyd, T. N. Woods, R. Cebula, E. Hilsenrath, M. Herse, and D. Labs, in Solar Variability and Its Effects on Climate. J. M. Pap, P. Fox, C. Frohlich, H. S. Hudson, J. Kuhn, J. McCormack, G. North, W. Sprigg and S. T. Wu, (eds.) (American Geophysical Union, 2004) pp. 171.

Maopeng Xia, Jianjun Li, Zhengqiang Li, Dongyang Gao, Weiwei Pang, Donghui Li, Xiaobing Zheng. Research on calibration method in lab of direct solar channels of Sun photometer[J]. Chinese Optics Letters, 2014, 12(12): 121201.