Chinese Optics Letters, 2016, 14 (6): 061201, Published Online: Aug. 3, 2018   

Novel homodyne frequency-shifting interference pattern locking system Download: 1041次

Author Affiliations
State Key Laboratory of Tribology, Department of Mechanical Engineering, Tsinghua University, Beijing 100084, China
Figures & Tables

Fig. 1. Schematic diagram of the novel homodyne frequency-shifting interference pattern locking system.

下载图片 查看原文

Fig. 2. Detailed scheme of the homodyne redundant phase-measurement interferometer with four interference signals.

下载图片 查看原文

Fig. 3. Photograph of experiment setup: (a) optics subsystem and (b) VME-based control subsystem.

下载图片 查看原文

Fig. 4. Plots of uncorrected and corrected experiment data sets: (a) Lissajous curve of signals from the HRPMI, (b) value of the HRPMI, (c) discrepancy, and (d) ASD for the data from (c).

下载图片 查看原文

Fig. 5. Experiment results: (a) interference pattern drift without locking, (b) residual drift with locking, and (c) ASD (in logarithmic coordinates) for the data from (a) and (b).

下载图片 查看原文

Fig. 6. Experiment results: (a) locking result and (b) error.

下载图片 查看原文

Table1. Short-Time Error of the Proposed System

Error CategoryValue (rad) (3σ)
HRPMI electronics error0.0024
HRPMI corrected error0.0024
Environmental error0.0132
Control error0.0461
Total error (RSS)0.0481

查看原文

Yu Zhu, Leijie Wang, Ming Zhang, Sen Lu, Kaiming Yang, Chuxiong Hu. Novel homodyne frequency-shifting interference pattern locking system[J]. Chinese Optics Letters, 2016, 14(6): 061201.

本文已被 5 篇论文引用
被引统计数据来源于中国光学期刊网