应用光学, 2017, 38 (3): 463, 网络出版: 2017-06-30   

高光谱遥感器现场光谱定标精度验证方法研究

Verification method of spectral calibration accuracy for hyperspectral remote sensors
作者单位
1 中国科学院 通用光学定标与表征技术重点实验室, 安徽 合肥 230031
2 中国科学院 安徽光学精密机械研究所, 安徽 合肥 230031
摘要
鉴于精确的光谱辐射定标精度验证对于高光谱遥感器的现场定标和数据应用非常必要, 以SVC光谱辐亮度计为例, 利用多波段辐亮度标准传递探测器和新型的光谱可调积分球参考光源, 设计了一种高光谱遥感器光谱定标精度的验证方法。该方法利用新型光谱可调积分球参考光源在待测波段内分别输出光谱形状单调上升和单调下降的光谱辐亮度状态, 通过光谱匹配技术, 即平移改变SVC光谱辐亮度计的波长, 分析比对MRSTD和SVC光谱辐亮度计测量辐亮度的相对偏差。比对结果为光谱定标验证精度优于±0.2 nm, 辐射定标验证精度小于5%。
Abstract
Validation for radiometric and spectral calibration accuracy of hyperspectral remote sensor is necessary. Taking SVC spectral radiometer as an example, new type spectrum-tunable reference light source (STIS) and muti-band radiance standard transfer detector(MRSTD) are used to design validation method for radiometric and spectral calibration accuracy. In this method, spectral level of spectral shape is monotonically increasing and monotonically decreasing in measured band by new spectral adjustable integrating sphere reference light source. Wavelength of SVC spectral radiance is changed by spectral matching technique. Relative deviation of radiance measured by MRSTD and SVC spectral radiometer are calculated and contrasted. Contrast shows that spectral calibration accuracy is better than ±0.2 nm, and absolute radiometric calibration accuracy is less than 5%.
参考文献

[1] 翟文超, 徐骏, 郑小兵, 等. 应用于超光谱定标的光谱可调光源光学设计[J].红外与激光工程, 2014,43(3): 951-955.

    Zhai Wenchao, Xu Jun, Zheng Xiaobing, et al. Optical design of a spectrally tunable source for hyperspectral calibration [J]. Infrared and Laser Engineering, 2014,43(3): 951-955.

[2] 乔延利, 郑小兵, 王先华,等. 卫星光学传感器全过程辐射定标[J]. 遥感学报, 2006, 10(5): 616-623.

    Qiao Yanli, Zheng Xiaobing, Wang Xianhua, et al. Whole-process radiometric calibration of optical remote sensors [J]. Journal of Remote Sensing, 2006, 10(5): 616-623.

[3] 郑小兵. 高精度卫星光学遥感器辐射定标技术[J].航天返回与遥感, 2011, 32 (05): 36-43.

    Zheng Xiaobing. High-accuracy radiometric calibration of satellite optical remote sensors [J]. Spacecraft Recovery & Remote Sensing, 2011, 32 (05): 36-43.

[4] 李新,张国伟,寻丽娜,等. 短波红外平场光谱仪的波长定标[J].光学学报, 2008, 28 (5): 920-926

    Li Xin, Zhang Guowei, Xun Lina, et al. Wavelength calibration of shortwave infrared flat spectroradiometer[J]. Acta Optica Sinica, 2008, 28 (5): 920-926.

[5] 徐秋云,郑小兵,李健军,等. 外部导入激光的积分球辐射源的研制[J]. 光学精密工程,2009, 17(4): 738-744.

    Xu Qiuyun, Zheng Xiaobing, Li Jianjun, et al.. Development of laser illuminated integrating sphere source [J]. Optics and Precision Engineering, 2009, 17(4): 738-744.

[6] 陈洪耀,张黎明,李鑫,等.高光谱遥感器飞行中基于大气特征谱线的光谱定标技术[J].光学学报, 2013, 33 (5): 287-293.

    Chen Hongyao, Zhang Liming, Li Xin, et al. Hyperspectral sensor in flight spectral calibration based on characteristic spectra of atmosphere [J]. Actc Optica Sinica, 2013, 33 (5): 287-293.

[7] Green R O. Spectral calibration requirement for Earth-looking imaging spectrometers in the solar-reflected spectrum[J]. Applied Optics, 1998, 37(4): 683- 690.

[8] Gao Bocai, Montes M J, Davis C O.Refinement of wavelength calibrations of hyperspectral imaging datausing a spectrum-matching technique[J] . Remote Sensing of Environment, 2004, 90(4): 424-433.

[9] Green R O. Determination of the in-flight spectral calibration of AVIRIS using atmospheric absorption features[M].US: JPL Publication, 1995.

[10] 陈伟,晏磊,勾志阳,等.无人机多光谱传感器场地绝对辐射定标研究[J].光谱学与光谱分析,2012, 32(12): 3169-3174.

    Chen Wei, Yan Lei, Gou Zhiyang, et al. In-flight absolute radiometric calibration of UAV multispectral sensor [J]. Spectroscopy and Spectral Analysis, 2012, 32(12): 3169-3174.

[11] 勾志阳,晏磊,陈伟,等.无人机高光谱成像仪场地绝对辐射定标及验证分析[J].光谱学与光谱分析,2012,32(5): 430-434.

    Gou Zhiyang, Yan Lei, Chen Wei, et al. In-flight absolute radiometric calibration of UAV hyperspectral camera and its validation analysis[J]. Spectroscopy and Spectral Analysis, 2012, 32(5): 430-434.

[12] 袁银麟,徐骏,翟文超,等. 大孔径可调光谱积分球参考光源研制和检测[J]. 光学学报, 2013, 33(7): 0712004-1-8.

    Yuan Yinlin, Xu Jun, Zhai Wenchao, et al. Design and test of a spectrally tunable integrating sphere reference light source with large exit aperture [J]. Acta Optica Sinca, 2013, 33(7): 0712004-1-8.

[13] 张艳娜,郑小兵,李健军,等.多波段滤光片辐亮度计的系统级定标方法研究[J].光学学报, 2013, 33(8): 0812004-1-7.

    Zhang Yanna, Zheng Xiaobing, Li Jianjun, et al.Research on system level calibration method of multi-band filter radiometer [J]. Acta Optica Sinca,2013, 33(8): 0812004-1-7.

[14] Brown S W, Johnson B C, Feinholz M E, et al. Stray-light correction algorithm for spectrographs[J]. Metrologia, 2003m 40(1)81-84.

[15] 倪育才.实用测量不确定度评定[M].第三版.北京: 中国计量出版社, 2009.

    Ni Yucai. Evaluation of practical measurement uncertainty[M].3rd ed.Beijing: Chinese Metrology Press, 2009.

丁蕾, 袁银麟, 郑小兵, 张艳娜, 戚涛, 吴浩宇. 高光谱遥感器现场光谱定标精度验证方法研究[J]. 应用光学, 2017, 38(3): 463. Ding Lei, Yuan Yinlin, Zheng Xiaobing, Zhang Yanna, Qi Tao, Wu Haoyu. Verification method of spectral calibration accuracy for hyperspectral remote sensors[J]. Journal of Applied Optics, 2017, 38(3): 463.

本文已被 2 篇论文引用