激光与光电子学进展, 2013, 50 (12): 123002, 网络出版: 2013-11-13   

土壤中铜的单脉冲LIBS定标方法研究

Research on the Method of Calibration of Cu in Soil by Single Pulse Laser-Induced Breakdown Spectroscopy
作者单位
1 西南科技大学信息工程学院, 四川 绵阳 621010
2 西南科技大学极端条件物质特性实验室, 四川 绵阳 621010
摘要
定标法对激光诱导击穿光谱(LIBS)技术的定量分析精度有着重要影响。在利用激光诱导击穿光谱技术探测土壤样品中铜元素的过程中,采用外标法与不同的内标法定标曲线对土壤样品中的铜元素进行了定量分析。在以Ca作为内标元素的分析中,通过Voigt拟合对分析谱线与内标元素谱线轮廓进行预处理,提取光谱净强度值,然后根据分析线与内标线净强度之比建立定标曲线,所得的线性系数为0.998,通过循环反演得到的平均相对偏差为2.28%且相对偏差波动很小,具有不错的稳定性。以Cu 324.754 nm的峰值强度与谱图在323~326 nm处的面积之比建立定标曲线,得到定标曲线的线性系数为0.997,通过循环反演得到的平均相对偏差为2.23%,稳定性略低于前一种内标法。采用外标法直接用样品中Cu浓度与Cu 324.754 nm处峰值强度建立的定标曲线的线性系数为0.972,通过循环反演得到的平均相对偏差为13%且波动较大,可靠性较差。
Abstract
Calibration method has an important impact on the accuracy of quantitative analysis of laser-induced breakdown spectroscopy (LIBS). In the process of using LIBS to detect copper in soil samples, different calibration curves are used in order to do quantitative analysis, including exterior standard method and different internal standard methods. Using Voigt fitting function to the spectral line at Cu 324.754 nm and Ca 317.9 nm to obtain the net intensity values, the net intensity ratio between Cu and Ca is then sought as internal standard to get calibration curve, which has the correlation coefficient up to 0.998. Through the cyclic inversion, the average relative error is 2.28% and the volatility is very small, so this method has good stability. Using the ratio between the spectra line intensity of Cu 324.754 nm and the area from 323 nm to 326 nm to establish the calibration curve, the correlation coefficient is 0.997 and the average relative error by cyclic inversion is 2.23%, but the stability is slightly lower than the former one. Directly using the Cu concentration and spectra line intensity to establish the calibration curve, the correlation coefficient is 0.972 and the average relative error by cyclic inversion is 13%, showing large fluctuations and less reliability.

李文宏, 武志翔, 王芮雯, 尚丽平. 土壤中铜的单脉冲LIBS定标方法研究[J]. 激光与光电子学进展, 2013, 50(12): 123002. Li Wenhong, Wu Zhixiang, Wang Ruiwen, Shang Liping. Research on the Method of Calibration of Cu in Soil by Single Pulse Laser-Induced Breakdown Spectroscopy[J]. Laser & Optoelectronics Progress, 2013, 50(12): 123002.

本文已被 3 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!