中国激光, 2006, 33 (6): 734, 网络出版: 2006-06-13   

拉曼激光雷达信号采集及处理系统

Scheme on Acquisition and Disposal for Raman Lidar System
作者单位
1 中国科学院安徽光学精密机械研究所大气光学重点实验室, 安徽 合肥 230031
2 安徽省环境光学监测技术重点实验室, 安徽 合肥 230031
摘要
准确可靠地监测大气中CO2含量变化对于研究全球变暖和气候变化具有十分重要的意义。利用拉曼激光雷达测量大气中的CO2含量分布是一种新颖且容易实现的方法。有效地采集CO2气体的拉曼回波信号并进行数据处理反演出测量结果是整个系统的关键所在。根据拉曼激光雷达的原理和信号处理的理论基础,设计了单通道和双通道两种行之有效的拉曼信号采集方案,并分别对其信号处理方法进行了论证。阐述了单通道采集过程中能量波动所造成的对信号的影响和消除方法,重点解决了采集过程中存在的噪声干扰问题。计算了两种方案在采集1 km处的系统信噪比(SNR),单通道采集时约可达到10,而双通道采集时可提高到20左右。
Abstract
Detecting the atmospheric CO2 reliably is becoming more and more important due to environment problems and climate changing. Raman light detection and range (lidar) system monitoring the atmosphere is a novel technique. It is a key factor to acquire the back signals of Raman lidar efficiently. The paper not only analyzes the principle of Raman lidar and the theoretical basis of signal processing, but also designs two schemes for collecting Raman signals, which include single channel and double channels, and demonstrates the methods of signal processing separately. The paper resolves the noise interference in collecting signals process, and gets the satisfied conclusion by contrasting the two results of the two schemes. Moreover, the signal-to-noise ratio (SNR) is calculated for the lidar system. The calculated results indicate that the SNR (for 1 km) is up to 10 for single channel, however, up to about 20 for double channels.
参考文献

[1] Michihiro Uchiumi, Nilesh Jvasa, Makoto Fujiwara et al.. Development of DIAL for CO2 and CH4 in the atmosphere [C]. SPIE, 2003, 4893:141~149

[2] . Refaat, M. Nurul Abedin, Upendra N. Singh. Spectral response measurements of short wave infrared detectors (SWIR)[J]. International Journal of High Speed Electronics and Systems, 2002, 12(2): 541-550.

[3] Wu Yonghua, Yue Guming, Hu Huanling et al.. D2 stimulated Raman scattering pumped by fourth harmonic Nd:YAG laser and its application in laser radar [J]. Chinese J. Lasers, 2000, A27(9):823~827
吴永华,岳古明,胡欢陵 等. Nd:YAG四倍频激光抽运氘后的拉曼效应及其在激光雷达中的应用[J]. 中国激光, 2000, A27(9):823~827

[4] . Park. Atmospheric CO2 monitoring from space[J]. Appl. Opt., 1997, 36(12): 2701-2712.

[5] Wu Yonghua, Hu Huanling, Hu Shunxing et al.. Rayleigh-Raman scattering lidar for atmospheric temperature profiles measurements [J]. Chinese J. Lasers, 2004, 31(7):851~856
吴永华,胡欢陵,胡顺星 等. 瑞利-拉曼散射激光雷达探测大气温度分布[J]. 中国激光, 2004, 31(7):851~856

[6] . . Experimental investigation of high-power single-pass Raman shifters in the ultraviolet with Nd:YAG and KrF lasers[J]. Appl. Opt., 1997, 36(21): 5026-5043.

[7] . Comparison among error calculations in differential absorption lidar measurements[J]. Optics & Laser Technology, 2001, 33(6): 371-377.

[8] Tao Zongming, Zhang Yinchao, Cen Gang et al.. Estimating method of detecting minimum pollutant gas concentration by lidar [J]. Acta Optica Sinica, 2004, 24(5):602~604
陶宗明,张寅超,岑岗 等. 激光雷达探测污染气体最小浓度的估算方法[J]. 光学学报, 2004, 24(5):602~604

[9] Hu Shunxing, Hu Huanling, Zhang Yinchao et al.. Differential absorption lidar for environmental SO2 measurement [J]. Chinese J. Lasers, 2004, 31(9):1121~1126
胡顺星,胡欢陵,张寅超 等. 差分吸收激光雷达测量环境SO2[J]. 中国激光, 2004, 31(9):1121~1126

赵曰峰, 张寅超, 洪光烈, 赵培涛, 苏嘉, 方欣, 谢军, 屈凯峰. 拉曼激光雷达信号采集及处理系统[J]. 中国激光, 2006, 33(6): 734. 赵曰峰, 张寅超, 洪光烈, 赵培涛, 苏嘉, 方欣, 谢军, 屈凯峰. Scheme on Acquisition and Disposal for Raman Lidar System[J]. Chinese Journal of Lasers, 2006, 33(6): 734.

本文已被 6 篇论文引用