应用光学, 2018, 39 (6): 862, 网络出版: 2019-01-10  

滤光辐射计测量黑体的温度

Temperature measurement of blackbody using filter radiometer
作者单位
西安应用光学研究所, 陕西 西安 710065
摘要
介绍了黑体温度测量的溯源途径及方法, 分析了探测器绝对光谱响应、透镜透过率及面积效应等对黑体温度测量的影响, 试验测量结果表明黑体温度可以通过标准探测器进行量值传递并溯源。与传统溯源方法相比, 利用标准探测器进行量值传递测量精度可提升一个数量级, 达到0.8%的水平。
Abstract
The tracing ways and methods for temperature measurement of blackbody were introduced. The influences of the absolute spectral response of detector,the lens transmittance and the area effect on the blackbody temperature measurement were analyzed.The test results show that the blackbody temperature can be transmitted and traced by the standard detector. Compared with traditional tracing methods, the accuracy of blackbody temperature measurement transmitted by the standard detector can increased by an order of magnitude, reaching a level of 0.8%.
参考文献

[1] 郝小鹏,宋健,孙建平,等. 风云卫星的红外遥感亮度温度国家计量标准装置[J]. 光学精密工程, 2015, 23(7): 1845-1851.

    HAO Xiaopeng , SONG Jian , SUN Jianping , et al.Vacuum radiance temperature national standard facility for infrared remote sensors of chinese fengyun meteorological satellites[J]. Optics and Precision Engineering, 2015, 23(7): 1845-1851.

[2] 徐骏,孟炳寰,翟文超,等. 基于热红外标准辐亮度计的常温黑体定标技术[J]. 红外与激光工程, 2014, 43(3): 716-721.

    XU Jun, MENG Binghuan, ZHAI Wenchao. Calibration of common temperature blackbody based on thermal-infrared standard radiometer[J].Infrared and laser Engineering,2014, 43(3): 716-721.

[3] 郑峰,刘丽莹,刘小溪,等. 基于中温黑体的近红外光纤光谱仪辐射定标的方法研究[J]. 光谱学与光谱分析, 2015, 35(9): 2665-2670.

    ZHENG Feng,LIU Liying,LIU Xiaoxi,et al.VMTBB-Based spectral radiometric calibration of NIR fiber coupled spectrometer[J].Spectroscopy and Spectral Analysis, 2015, 35(9): 2665-2670.

[4] 李云红,孙晓刚,原桂彬,等. 红外热像仪精确测温技术[J]. 光学精密工程, 2007, 15(9): 1336-1341.

    LI Yunhong, SUN Xiao gang, YUAN Guibin,et al.Accurate measuring temperature with infrared thermalimager[J]. Optics and Precision Engineering, 2007, 15(9): 1336-1341.

[5] 张艳娜,郑小兵,李健军,等. 多波段滤光片辐亮度计的系统级定标方法研究[J].光学学报, 2013, 33(8): 123-129.

    ZHANG Yanna,ZHENG Xiaobing,LI Jianjun, et al. Research on system level calibration method of multi-band filter radiometer[J]. Acta Optica Sinica, 2013, 33(8): 123-129.

[6] 杨超普, 方文卿, 刘明宝,等. MOCVD原位红外测温方法的比较研究[J]. 应用光学, 2016, 37(2)297-302.

    YANG Chaopu,FANG Wenqing,LIU Mingbao, et al. Comparative study on in situ infrared thermometry methods of MOCVD[J]. Journal of Applied Optics, 2016, 37(2): 297-302.

[7] 周书铨. 红外辐射测量基础[M].上海: 上海交通大学出版社, 1991: 11-20.

    ZHOU Shuquan. Infrared radiation measurement base[M].Shanghai: Shanghai Jiao Tong University Press, 1991: 11-20.

[8] YOON HW,ALLEN DW,GIBSON CE, et al. Thermodynamic-temperature determinations of the ag and au freezing temperatures using a detector-based radiation thermometer[J]. Applied Optics, 2007, 46(15): 2870-80.

[9] 辛成运. 温度场辐射测量及其标定研究[D].合肥: 中国科学技术大学, 2013.

    XIN Chengyun. Research of temperature field radiation measurement and its calibration[D].Hefei: University of Science and Technology of China, 2013.

[10] 范纪红,侯西旗,杨照金,等. 红外探测器光谱响应度测试技术研究[J]. 应用光学, 2006, 27(5): 460-462.

    FAN Jihong,HOU Xiqi,YANG Zhaojin, et al. Measurement technology for spectral responsivity of infrared detector[J]. Journal of Applied Optics, 2006, 27(5): 460-462.

[11] YOON HW,EPPELDAUER GP. Measurement of thermal radiation using regular glass optics and short-wave infrared detectors[J]. Optics Express, 2008, 16(2): 937-49.

[12] WOOLLIAMS E R, POLLARD D P, HARRISON N J E. T and Fox N P. new facility for the high-accuracy measurement of lens transmission[J]. Metrologia,2003, 37(5): 603-605.

[13] BLOEMBERGEN P. Analytical representations of the size-of-source effect[J]. Metrologia , 2009 , 46: 534-543。

[14] BART M, HAM EWMVD, SAUNDERS P. A new method to determine the size-of-source effect[J]. International Journal of Thermophysics,2007, 28(6): 2111-2117.

于东钰, 张魁甲, 吴遥, 朱银生, 吴沛, 阴万宏, 李四维, 岳文龙. 滤光辐射计测量黑体的温度[J]. 应用光学, 2018, 39(6): 862. Yu Dongyu, Zhang Kuijia, Wu Yao, Zhu Yinsheng, Wu Pei, Yin Wanhong, Li Siwei, Yue Wenlong. Temperature measurement of blackbody using filter radiometer[J]. Journal of Applied Optics, 2018, 39(6): 862.