光学学报, 2010, 30 (4): 1026, 网络出版: 2010-04-20   

真空紫外探测器辐射定标研究

Investigation on Radiation Calibration System of Vacuum Ultraviolet Detector
作者单位
1 中国科学院 长春光学精密机械与物理研究所 应用光学国家重点实验室,吉林 长春 130033
2 中国科学院研究生院,北京 100049
摘要
为了满足紫外光学遥感信息定量化发展的要求,构建了一套高精度真空紫外探测器辐射定标系统,以美国国家标准技术研究院(NIST)真空紫外标准探测器为标准,在120-220 nm波段上,对量子效率未知的EMR541-09-17紫外光电管进行了定标。按照国际通用不确定度评估规范,对测量结果进行不确定度分析和评估,总的传递不确定度小于2.4%。达到了在真空紫外波段对紫外探测器进行高精度定标的要求。
Abstract
A high-accuracy radiation calibration system of vacuum ultraviolet (VUV) detector is established,to meet the development of the ultraviolet (UV) and VUV remote quantification. In 120-220 nm spectral range,the absolute responsivity was transferred to EMR541-09-17 ultraviolet phototube detector,on the basic of national institute of standards and technology (NIST) ultraviolet standard detector. According to the guidelines for evaluating and expressing the uncertainty of NIST measurement,the relative uncertainty of transferring is less than 2.4%. The aim of high accuracy radiation calibration of ultraviolet detector in VUV region has been obtained.
参考文献

[1] 王淑荣,邢进,李福田. 利用积分球光源定标空间紫外遥感光谱辐射计[J]. 光学精密工程,2006,14(2):186-190

    Wang Shurong,Xing Jin,Li Futian. Spectral radiance responsivity calibration of ultraviolet remote sensing spectroradiometer in space using integrating sphere[J]. Opt. Precision Eng.,2006,14(2):186-190

[2] 邢进,王淑荣,李福田. 空间紫外遥感光谱辐射计光谱辐亮度定标三种方法的比较[J]. 中国激光,2006,33(4):509-515

    Xing Jin,Wang Shurong,Li Futian. Comparison of spectral radiance calibrations of spectroradiometer for ultraviolet space remote sensing using three calibration techniques[J]. Chinese J. Lasers,2006,33(4):509-515

[3] . A. R. Samson. Absolute intensity measurements in the vacuum ultraviolet[J]. J. Opt. Soc. Am, 1964, 54(1): 6-14.

[4] 曹光宇,张志伟,张存林. 光电检测技术[M].北京:清华大学出版社,2005

    Cao Guagyu,Zhang Zhiwei,Zhang Cunlin. Technology of Photoelectricity Testing[M]. Beijing:Tsinghua University Press,2005

[5] . G. Johnston,R.P. Madden. On the use of thermopiles for absolute radiometry in the far ultraviolet[J]. Appl. Opt., 1965, 4(12): 1574-1580.

[6] . F. Zalewski ,J. Geist. Silicon photodiode absolute spectral response self-calibration[J]. Appl. Opt., 1980, 19(8): 1214-1216.

[7] E. F. Zalewski. The NBS Photodetector Spectral Response Calibration Transfer Program,Natl[M]. Bur. Stand.,Spec. Publ.,1988

[8] 费业泰.误差理论与精度分析[M].北京:国防工业出版社,2004

    Fei Yetai. Theory of Error and Precision and Analysis[M].Beijing:National Defense Industry Press,2004

[9] 王淑荣,宋克非,李福田. 星载太阳紫外光谱监视器的地面辐射定标[J]. 光学学报,2007,27(12):2256-2261

    Wang Shurong,Song Kefei,Li Futian. Radiometric calibration on ground of satellite-borne solar ultraviolet spectral monitor[J]. Acta Optica Sinica,2007,27(12):2256-2261

王锐, 王淑荣, 李福田, 刘海波, 汪龙祺, 黄煜. 真空紫外探测器辐射定标研究[J]. 光学学报, 2010, 30(4): 1026. Wang Rui, Wang Shurong, Li Futian, Liu Haibo, Wang Longqi, Huang Yu. Investigation on Radiation Calibration System of Vacuum Ultraviolet Detector[J]. Acta Optica Sinica, 2010, 30(4): 1026.

本文已被 7 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!