光学学报, 2008, 28 (11): 2159, 网络出版: 2008-11-17   

一种现场大尺寸测量精度的评价方法

A Method of Precision Evaluation for Field Large-Scale Measurement
作者单位
天津大学精密测试技术及仪器国家重点实验室, 天津 300072
摘要
针对现场全面评价大尺寸测量仪器精度的特殊性, 提出了一种采用多仪器站多控制点的精度评价方法。基于四元数的空间数据配准, 将多仪器站的测量数据统一到全局坐标系下, 基于控制点坐标不变约束, 对测量值进行统计得到测量不确定度, 并从中提取仪器内部各传感器单元分量的不确定度。利用多传感器信息融合技术求得最优控制点, 减小坐标转换误差对结果的影响。以激光跟踪仪为例进行了仿真和现场实验, 结果表明, 该方法评价测距和测角不确定度的误差可分别降至1 μm和0.1″以内。
Abstract
A method of precision evaluation for field large-scale measurement instruments is proposed with multi-instrument stations and multi-reference points. The measured data from multi-instrument stations are unified in the global coordinate system using spatial data registration based on quaternion. The measurement uncertainty is obtained by statistics of measurement values based on controlling points coordinate invariable constraint, so that the uncertainty of each sensor component is acquired. The optimal controlling points are calculated by multi-sensors information fusion technique, and the effect of coordinate transform error is decreased. Taking laser tracker as an example, the simulation and field experiments show that the uncertainty error of range and angle are decreased to 1 μm and 0.1″, respectively.
参考文献

[1] . T. Estler, K. L. Edmundson, G. N. Peggs et al.. Large-scale metrology-an update[J]. CIRP Annals-Manufacturing Technology, 2002, 51(2): 587-609.

[2] . . Large 3D free surface measurement using a mobile coded light-based stereo vision system[J]. Sensors and Actuators A, 2006, 132(2): 460-471.

[3] J. M. Calkins. Quantifying coordinate uncertainty fields in coupled spatial measurement systems[D]. Virginia Polytechnic Institute, Doctor of Philosophy, 2002

[4] 雷琳君, 杨燕, 陈卫标. 星载激光测距仪的高精度时间间隔测量单元[J]. 中国激光, 2007, 34(10): 1422~1426

    Lei Linjun, Yang Yan, Chen Weibiao. High-accuracy time-interval-unit for spaceborne laser range finder[J]. Chin. J. Lasers, 2007, 34(10): 1422~1426

[5] James H. Burge, Peng Su, Chunyu Zhao et al.. Use of a commercial laser tracker for optical alignment[C]. Proc. SPIE, 2007, 6676: 1~12

[6] 吴庆阳, 苏显渝, 向立群 等. 线结构光双传感器测量系统的标定方法[J]. 中国激光, 2007, 34(2): 259~264

    Wu Qingyang, Su Xianyu, Xiang Liqun et al.. A new calibration method for two-sensor measurement system based on line-structure light[J]. Chin. J. Lasers, 2007, 34(2): 259~264

[7] 敖磊, 谭久彬, 崔继文 等. 一种快速高精度激光CCD自准直仪圆目标中心的定位方法[J]. 光学学报, 2007, 27(2): 253~258

    Ao Lei, Tan Jiubin, Cui Jiwen et al.. Fast and precise center location for circle target of CCD laser autocollimator[J]. Acta Optica Sinica, 2007, 27(2): 253~258

[8] T. A. Clarke, X. Wang, N. R. Cross et al.. Performance verification for large volume metrology systems[J]. Laser Metrology and Machine Performance V, WIT Press. 2001. 105~116

[9] . 动态加权在激光跟踪仪球坐标整合中的应用[J]. 哈尔滨工业大学学报, 2007, 39(9): 1419-1421.

    . . Application of dynamic weighting in HVD coordinate integrated algorithm[J]. J. Harbin Institute of Technology, 2007, 39(9): 1419-1421.

[10] 丁雅斌, 彭翔, 田劲东 等. 一种三维数字成像系统的多视点姿态估计方法[J]. 光学学报, 2007, 27(3): 451~456

    Ding Yabin, Peng Xiang, Tian Jindong et al.. Pose estimation of multiple viewpoints for three-dimensional digital imaging system[J]. Acta Optica Sinica, 2007, 27(3): 451~456

张福民, 曲兴华, 戴建芳, 叶声华. 一种现场大尺寸测量精度的评价方法[J]. 光学学报, 2008, 28(11): 2159. Zhang Fumin, Qu Xinghua, Dai Jianfang, Ye Shenghua. A Method of Precision Evaluation for Field Large-Scale Measurement[J]. Acta Optica Sinica, 2008, 28(11): 2159.

本文已被 4 篇论文引用