抑制容器荧光干扰的双轴共焦拉曼检测方法 下载: 827次
Aimed at eliminating the fluorescence interference from containers of the sample in the detection of Raman spectroscopy, a dual-axis confocal detection scheme is proposed based on the characteristic that the fluorescence generated by the container and the Raman signal of the sample are not confocal. The experimental results show that the fluorescence interference signals from the container are reduced by an order of magnitude,and the intensity of Raman signal decreases by about 30%. The dual-axis confocal Raman detection scheme can solve the mixed reception problem of Raman signal and fluorescence in the conventional coaxial confocal Raman detection scheme, solve the limitation of the dynamic detection range of Raman signal caused by the fluorescence interference, and achieve the effective detection of Raman signal of the material in fluorescent material containers.
1 引言
拉曼光谱检测技术具有简便快捷、灵敏度高和无损检测等优点,在生物医学、化学分析、食品安全和考古等领域应用广泛[1-4]。目前,荧光干扰是阻碍拉曼光谱检测技术发展的一个普遍问题。当受到激光照射时,被测样品不仅会被激发出表征信息的拉曼散射光,还有可能被激发出荧光,如样品装载容器的荧光和样品本身的荧光。通常情况下,荧光信号的强度远远大于拉曼散射光强,有的可以达到拉曼光强的上千倍[5-6],甚至会淹没拉曼信号,使拉曼光谱不可识别,因此研究荧光干扰的抑制方法十分必要,这对于拉曼光谱检测技术的发展具有重要意义。
随着拉曼光谱检测技术的发展,消除荧光干扰的方法层出不穷,主要有扣除荧光基线的软件算法[7-9]、差分拉曼法[10-12]和空间偏移拉曼光谱(SORS)技术[13-14]等。
扣除荧光基线软件算法和差分拉曼法都是对测量后的拉曼光谱数据进行处理的,在荧光干扰背景中提取拉曼信号。这些方法通常适用于处理荧光光谱变化比较平缓和拉曼信号比较强的情况,但由于算法误差的影响,这些方法难以识别弱小的拉曼信号。对于比较复杂的荧光光谱形状和噪声较大的光谱,这些方法的误差更大,甚至可能会产生误判。对于完全淹没在强荧光信号中的拉曼信号,这些方法更是无能为力。
当被测样品自身没有荧光或有弱荧光时,就需要解决容器产生的荧光干扰。SORS技术利用拉曼散射光在样品深层扩散时横向转移的特性,将检测点设置在偏移激发点一定距离处,这样虽然可以避免容器的荧光干扰,但接收到的拉曼散射光强度也变得很弱,对光谱探测器的灵敏度要求很高,导致硬件成本昂贵,不利于大范围推广使用。
采用无荧光效应的石英材料制作样品容器也是一种消除荧光的解决方案,但石英材料价格是普通玻璃材料(如硼材料)的几十倍,增加了测试成本,降低了产品竞争力。
目前,大多数拉曼设备的检测探头都采用共轴共焦光学结构,这种探头的激光发射和拉曼信号接收光路是共轴重合的,在接收拉曼散射光的同时无法避免荧光信号的接收。本文提出一种抑制样品装载容器荧光接收的双轴共焦拉曼检测方法,激发光路和拉曼信号接收光路的光轴在空间上有一定夹角,通过聚焦透镜共焦在一点,与共轴共焦光学结构一样满足物(发射光纤)像(接收光纤)共轭关系,可以接收到在焦点处样品被激发的拉曼信号,而荧光激发区域由于偏离了共焦点位置,因此荧光信号不会被接收。该方法可以有效抑制容器荧光的干扰,解决了荧光干扰对拉曼信号动态范围调整的限制问题,拓展了拉曼光谱检测技术的应用范围。
2 原理和方法
目前,拉曼光谱检测设备采用的拉曼探头大都是Dilor、InPhotonics等公司的产品,是一种双光纤共轴共焦结构[15],如
共轴共焦光路原理如
图 1. (a)传统共轴共焦拉曼检测方法;(b)共轴共焦光路原理
Fig. 1. (a) Scheme of conventional confocal and coaxial Raman detection; (b) principle of confocal and coaxial optical path
在实际应用中,采用共轴共焦光路结构的拉曼探头进行检测时,激发光透过容器聚焦在样品上,含有荧光的容器与被测样品在同一激发和探测光路内,虽然容器偏离焦点位置,如
为解决荧光信号对拉曼信号的影响,设计了一种双轴共焦拉曼检测方法,如
式中:
图 2. (a)双轴共焦拉曼检测光路原理;(b)细节图
Fig. 2. (a) Principle of dual-axis confocal optical path; (b) detail figure
3 实验设计
3.1 实验装置
实验装置主要由激发光路、接收光路、光谱仪、360°旋转平台、被测样品组成,如
激发光路、接收光路分别采用传统共轴共焦拉曼探头的激发和接收部分,其中激发光纤的纤芯直径为105 μm,接收光纤的纤芯直径为200 μm,两者的数值孔径均为0.22±0.01。光谱仪使用Maya2000(美国海洋光学公司),其分辨率为3 nm;旋转平台为中空带有角度刻度的360°旋转平台。
在测试操作过程中,接收探头固定在旋转平台上面的夹持器上,激发探头固定在旋转平台外侧的五维调整架上,被测样品安装固定在旋转平台的中心,调整容器的内表面(样品的被测点)位于旋转平台中心,通过五维调整架和夹持器调整激发光焦点、接收光焦点、样品被测点重合在旋转平台的旋转中心轴线上,并调整接收光和激发光的光轴构成的平面与旋转面平行。当旋转平台转动时,可带动接收探头一起转动,从而改变激发光与接收光之间的夹角。
实验的具体光路如
3.2 实验结果
3.2.1 双轴夹角对拉曼信号强度的影响
采用双轴共焦检测方案测试在不同双轴夹角
图 5. 900.30 nm处拉曼信号强度和双轴夹角之间的关系
Fig. 5. Relationship between Raman intensity at 900.30 nm and angle of dual-axis
图 6. (a)山梨酚的拉曼光谱;(b) 900.30 nm处放大图
Fig. 6. (a) Raman spectra of sorbitol; (b) enlarge image at 900.30 nm
实验测试结果表明,采用双轴共焦检测方案时,当双轴夹角小于一定值(44°以内)时,对接收到的拉曼信号强度影响不大。
采用同样的测试方法,对空玻璃比色皿进行了荧光信号测试,结果发现:在34°~44°测量范围内,双轴夹角为34°时的荧光信号最大;接收到的荧光信号随着双轴夹角增大而不断减弱。
3.2.2 双轴夹角对荧光信号强度的影响
采用对比实验测试双轴夹角对玻璃比色皿容器(高硼硅)荧光的抑制作用,被测样品为山梨酚粉末。
从
对双酚A粉末也进行同样的对比测试,实验结果如
4 分析与讨论
由(1)式可知,根据透镜的数值孔径
5 结论
双轴共焦拉曼检测方法利用样品装载容器产生的荧光和样品拉曼散射光不共焦的特性,从光学原理上避免了拉曼检测中荧光和拉曼散射光混合接收的问题,达到了有效消除来自样品装载容器荧光干扰的目的,解决了光谱仪动态范围因被荧光干扰信号限制而无法进行调整的问题。
从实验测试结果中可以看到,所提检测方案对样品自身的荧光没有特别有效的分离作用。目前有效的抑制和消除样品自身荧光的方法是时间门控法,而更便利和低成本的抑制样品荧光的方案有待进一步研究。
[1] 尉颖琪. 拉曼光谱技术及其在医学中的应用[J]. 科技创新导报, 2011( 33): 81.
尉颖琪. 拉曼光谱技术及其在医学中的应用[J]. 科技创新导报, 2011( 33): 81.
Wei YQ. Raman spectroscopy and its application in medicine[J]. Science and Technology Consulting Herald, 2011( 33): 81.
Wei YQ. Raman spectroscopy and its application in medicine[J]. Science and Technology Consulting Herald, 2011( 33): 81.
[2] 胡军, 胡继明. 拉曼光谱在分析化学中的应用进展[J]. 分析化学, 2000, 28(6): 764-771.
胡军, 胡继明. 拉曼光谱在分析化学中的应用进展[J]. 分析化学, 2000, 28(6): 764-771.
Hu J, Hu J M. Applications progress of Raman techniques in analytical chemistry[J]. Chinese Journal of Analytical Chemistry, 2000, 28(6): 764-771.
[3] 王吉有, 王闵, 刘玲, 等. 拉曼光谱在考古中的应用[J]. 光散射学报, 2006, 18(2): 130-133.
王吉有, 王闵, 刘玲, 等. 拉曼光谱在考古中的应用[J]. 光散射学报, 2006, 18(2): 130-133.
[4] 房晓倩, 彭彦昆, 李永玉, 等. 基于表面增强拉曼光谱快速定量检测碳酸饮料中苯甲酸钠的方法[J]. 光学学报, 2017, 37(9): 0930001.
房晓倩, 彭彦昆, 李永玉, 等. 基于表面增强拉曼光谱快速定量检测碳酸饮料中苯甲酸钠的方法[J]. 光学学报, 2017, 37(9): 0930001.
[6] 陈珊. 拉曼光谱背景扣除算法及其应用研究[D]. 长沙: 中南大学, 2011.
陈珊. 拉曼光谱背景扣除算法及其应用研究[D]. 长沙: 中南大学, 2011.
ChenS. Research on background deduction algorithm of Raman spectrum and its application[D]. Changsha: Central South University, 2011.
ChenS. Research on background deduction algorithm of Raman spectrum and its application[D]. Changsha: Central South University, 2011.
[7] 王欢, 王永志, 赵瑜, 等. 拉曼光谱中荧光抑制技术的研究新进展综述[J]. 光谱学与光谱分析, 2017, 37(7): 2050-2056.
王欢, 王永志, 赵瑜, 等. 拉曼光谱中荧光抑制技术的研究新进展综述[J]. 光谱学与光谱分析, 2017, 37(7): 2050-2056.
[8] 杨桂燕, 李路, 陈和, 等. 基于广义Whittaker平滑器的拉曼光谱基线校正方法[J]. 中国激光, 2015, 42(9): 0915003.
杨桂燕, 李路, 陈和, 等. 基于广义Whittaker平滑器的拉曼光谱基线校正方法[J]. 中国激光, 2015, 42(9): 0915003.
[9] 赵芳, 彭彦昆. 区间偏最小二乘法结合拉曼光谱测定猪肉皮下脂肪的碘值[J]. 中国激光, 2017, 44(11): 1111001.
赵芳, 彭彦昆. 区间偏最小二乘法结合拉曼光谱测定猪肉皮下脂肪的碘值[J]. 中国激光, 2017, 44(11): 1111001.
[10] 李帆. 移频拉曼激发法抑制荧光干扰关键技术研究[D]. 厦门: 厦门大学, 2014.
李帆. 移频拉曼激发法抑制荧光干扰关键技术研究[D]. 厦门: 厦门大学, 2014.
LiF. Key techniques for suppression of fluorescence interference by frequency shift Raman excitation[D]. Xiamen: Xiamen University, 2014.
LiF. Key techniques for suppression of fluorescence interference by frequency shift Raman excitation[D]. Xiamen: Xiamen University, 2014.
[11] 吴景林, 范贤光, 许英杰, 等. 一种多波长激光频移激发的拉曼光谱测量装置: 201510057824.7[P].2015-05-06.
吴景林, 范贤光, 许英杰, 等. 一种多波长激光频移激发的拉曼光谱测量装置: 201510057824.7[P].2015-05-06.
Wu JL, Fan XG, Xu YJ, et al. Raman spectrum measuring device excited by multi-wavelength laser frequency shift excitation: 201510057825[P]. 2015-05-06.
Wu JL, Fan XG, Xu YJ, et al. Raman spectrum measuring device excited by multi-wavelength laser frequency shift excitation: 201510057825[P]. 2015-05-06.
[12] 牟涛涛, 熊胜军, 赵喜, 等. 便携式双波长消荧光拉曼光谱检测系统: 201521113688.0[P].2016-09-28.
牟涛涛, 熊胜军, 赵喜, 等. 便携式双波长消荧光拉曼光谱检测系统: 201521113688.0[P].2016-09-28.
Mou TT, Xiong SJ, ZhaoX, et al. Portable dual-wavelength fluorescence Raman spectrometry detection system: 201521113688.0[P]. 2016-09-28.
Mou TT, Xiong SJ, ZhaoX, et al. Portable dual-wavelength fluorescence Raman spectrometry detection system: 201521113688.0[P]. 2016-09-28.
[15] 殷磊, 蔡圣闻, 姜晓冰, 等. 一种便携式可调节的拉曼探头: 201410061973.6[P].2014-05-07.
殷磊, 蔡圣闻, 姜晓冰, 等. 一种便携式可调节的拉曼探头: 201410061973.6[P].2014-05-07.
YinL, Cai SW, Jiang XB, et al. A portable adjustable Raman probe: 201410061973.6[P]. 2014-05-07.
YinL, Cai SW, Jiang XB, et al. A portable adjustable Raman probe: 201410061973.6[P]. 2014-05-07.
[16] 安岩. 手持式拉曼光谱仪的光机系统技术研究[D]. 长春: 中国科学院长春光学精密机械与物理研究所, 2014.
安岩. 手持式拉曼光谱仪的光机系统技术研究[D]. 长春: 中国科学院长春光学精密机械与物理研究所, 2014.
AnY. Research on the optical machine system of handheld Ramans pectrometer[D]. Changchun: Changchun Institute of Optics, Fine Mechanics and Physics,Chinese Academy of Sciences, 2014.
AnY. Research on the optical machine system of handheld Ramans pectrometer[D]. Changchun: Changchun Institute of Optics, Fine Mechanics and Physics,Chinese Academy of Sciences, 2014.
[17] 肖新民. 拉曼光谱仪的两种共焦显微术的对比[J]. 现代科学仪器, 2005( 3): 17- 20.
肖新民. 拉曼光谱仪的两种共焦显微术的对比[J]. 现代科学仪器, 2005( 3): 17- 20.
Xiao XM. Comparison between two different Raman confocal techniques[J]. Modern Scientific Instruments, 2005( 3): 17- 20.
Xiao XM. Comparison between two different Raman confocal techniques[J]. Modern Scientific Instruments, 2005( 3): 17- 20.
[18] 王雅杰. 激光共焦拉曼光谱仪测控技术研究[D]. 北京: 北京理工大学, 2016.
王雅杰. 激光共焦拉曼光谱仪测控技术研究[D]. 北京: 北京理工大学, 2016.
Wang YJ. Research on measurement and control technology of laser confocal Raman spectrometer[D]. Beijing: Beijing Institute of Technology, 2016.
Wang YJ. Research on measurement and control technology of laser confocal Raman spectrometer[D]. Beijing: Beijing Institute of Technology, 2016.
戴艳, 董作人, 刘铭晖, 辛国锋, 孙延光, 蔡海文. 抑制容器荧光干扰的双轴共焦拉曼检测方法[J]. 中国激光, 2018, 45(7): 0711001. Yan Dai, Zuoren Dong, Minghui Liu, Guofeng Xin, Yanguang Sun, Haiwen Cai. A Dual-Axis Confocal Raman Detection Method for Suppressing Fluorescence Interference from Containers[J]. Chinese Journal of Lasers, 2018, 45(7): 0711001.