应用光学, 2016, 37 (4): 573, 网络出版: 2016-08-29  

Solar-Blind滤光片可见光谱段光密度连续测量

Continuous measurement for optical density of visible band of solar blind UV filters
作者单位
中国科学院 长春光学精密机械与物理研究所, 吉林 长春 130000
摘要
为实现Solar-Blind(日盲紫外)滤光片光密度在可见光谱段的高精度连续测试, 搭建了以氙灯配合单色仪作为光源、使用单光子计数器作为探测器的测试系统, 并对该系统的误差进行分析。系统主要装置有氙灯与单色仪组合的光源, 标准衰减装置和单光子计数器; 以替代法为基础进行测试, 将由单光子计数器探测到的衰减后的计数值作为参考计数值替代光源的初始计数值, 通过运算得到滤光片的可见光谱段光密度; 该测试系统将滤光片光密度的测试动态范围扩展到0~-11OD, 实现了可见光谱段内进行连续光谱测量的功能, 且降低了系统的测试误差。与使用窄带LED作为光源的测试系统相比, 测试不确定度由2%降低至0.5%, 相对重复性误差由0.2% 降低至0.13%。
Abstract
To achieve the large dynamic range and high-precise test for the cutoff depth of the out-band of solar blind untraviolet(UV) filters, a system which not only used xenon lamp with a monochromator as the standard light, but also used photon counter as the detector was proposed. Additional, the error of the system was analyzed. The major equipments were the xenon lamp with a monochromator, the standard reflective neutral attenuator and the single photo counter. Based on the substitution method the measurement was carried out. The photo counter detected the number of the xenon lamp through the attenuator as the reference to substitute the initial photo number of xenon lamp, then the cutoff depth of filters could be acquired. Results show that the test system can extend the testing dynamic range of filter optical density to 0~-11OD, it realizes the continuous measurement the visible spectral bands at 1 nm intervals, and greatly reduces the test error of the system.Compared with the test system using narrow band LED as the light source, the test uncertainty reduces from 2% to 0.5%,and the relative repeatability error is from 0.2% down to 0.13%.
参考文献

[1] 刘晓科, 马君, 唐辉, 等. 一种日盲区紫外信号探测系统前端[J]. 探测与控制学报, 2007, 29(6): 41-44.

    Liu Xiaoke, Ma Jun, Tang Hui, et al. A front-end of ultraviolet signal detection system based on solar-blind region[J]. Journal of Detection&Control, 2007, 29(6): 41-44.

[2] 吴礼刚, 何文荣, 胡晋荪, 等. 日盲紫外-可见光双光谱照相机系统[J]. 光学精密工程, 2010, 18(7): 1529-1534.

    Wu Ligang, He Wenrong, Hu Jinsun, et al. Solar blind UV and visible dual-spectral camera[J]. Optics and Precision Engineering, 2010, 18(7): 1529-1534.

[3] 戴利波. 紫外成像技术在高压设备带电检测中的应用[J]. 电力系统自动化, 2003, 27(20): 97-98.

    Dai Libo. The application of UV imagery technology in the high-tension apparatus detection[J]. Autormation of Electric Power Systems, 2003, 27(20): 97-98.

[4] 王丽萍, 李春, 金春水. “日盲”紫外折反射全景光学系统设计[J]. 光学精密工程, 2011,19(7): 1503-1509.

    Wang Liping, Li Chun, Jin Chunshui. Design of catadioptric omnidirectional imaging system in solar blind UV[J]. Optics and Precision Engineering, 2011,19(7): 1503-1509.

[5] 周伟, 马妮, 吴晗平. 近地层紫外探测作用距离及其影响因素研究[J]. 红外技术, 2011,33(6): 357-371.

    Zhou Wei, Ma Ni, Wu Hanping. Operating rang of ultraviolet detection system in the surface layer and the study of its influence factors[J]. Infrared Technology, 2011,33(6): 357-371.

[6] 王正凤, 付秀华, 张静, 等. 日盲紫外告警系统中成像滤光片的研制[J]. 中国激光, 2011,38(12): 1207002.

    Wang Zhenfeng, Fu Xiuhua, Zhang Jing, et al. Research of imaging filter in solar blind ultraviolet alarm system[J]. Chinese Journal of lasers, 2011,38(12): 1207002.

[7] 刘伟峰, 谢永杰, 赵乐志. 天空背景辐亮度测量与研究 [J]. 应用光学, 2012,33(2): 351-354.

    Liu Weifeng, Xie Yongjie, Zhao Lezhi. Measurment and research on sky background radiance luminance[J]. Journal of Applied Optics,2012,33(2): 351-354.

[8] Laiho K, Christ A, Cassemiro K N, et al. Testing spectral filters as Gaussian quantum optical channels[J]. Optics Letters, 2011, 36(8): 1476-1478.

[9] David A, Compton C, John D, et al. Accurate infrared transmittance measurements on optical filters using an FT-IR spectrometer[J]. Applied Optics, 1990, 29(19): 2908-2912.

[10] 王锐, 王淑荣, 郭劲, 等. 高精度紫外标准探测器的定标[J]. 光学精密工程, 2012, 20(8): 1696-1703.

    Wang Rui, Wang Surong, Guo Jin, et al. Calibration of high accuracy UV standard detector[J]. Optics and Precision Engineering, 2012, 20(8): 1696-1703.

[11] 汪舰, 魏建平, 杨波, 等. 高透明液体透射率的测定[J]. 光学精密工程, 2013, 21(3): 624-629.

    Wang Jian, Wei Jianping, Yang Bo, et al. Measurement of transmittance for highly transparent liquid[J]. Optics and Precision Engineering, 2013, 21(3): 624-629.

[12] 陈栋, 黄云, 张斌, 等. 透反射率自动化测量系统设计[J]. 光学技术, 2007, 33(增刊): 210-213.

    Chen Dong, Huang Yun, Zhang Bin, et al. Design of transmissivity and reflectivity automatic measuring system[J]. Optical Technique, 2007,33(sup): 210-213.

[13] 范纪红, 赵生禄, 占春连, 等. 成像光谱仪绝对辐射标定技术研究 [J]. 应用光学, 2013, 34(4) : 210-213.

    Fan Jihong, Zhao Shenglu, Zhan Chunliang, et al. Absolute radiometric calibration technique of imaging spectrometer [J]. Journal of Applied Optics Optical,2013,34(4): 310-313.

[14] 刘卫国, 孙鑫, 高爱华, 等. 线性渐变滤光片透过率实时测量系统[J]. 西安工业大学学报, 2009,29(6): 515-518.

    Liu Weiguo,Sun Xin, Gao Aihua. Linear variable filters in real time transmittance measurement system[J]. Journal of Xi’an Technology University,2009,29(6): 515-518.

易翔宇, 周跃, 闫丰, 章明朝, 陈雪, 崔穆涵. Solar-Blind滤光片可见光谱段光密度连续测量[J]. 应用光学, 2016, 37(4): 573. Yi Xiangyu, Zhou Yue, Yan Feng, Zhang Mingchao, Chen Xue, Cui Muhan. Continuous measurement for optical density of visible band of solar blind UV filters[J]. Journal of Applied Optics, 2016, 37(4): 573.