光学学报, 2011, 31 (6): 0612008, 网络出版: 2011-05-31   

基于标准探测器研究标准灯光谱辐照度和漫反射板双向反射分布函数随波长的变化

Wavelength Dependence of the Spectral Irradiance of Standard Lamp and Bi-Reflection Distribution Function of Diffuser Based on Standard Detectors
作者单位
1 中国科学院长春光学精密机械与物理研究所, 吉林 长春 130033
2 中国科学院研究生院, 北京 100049
摘要
用光谱辐照度标准灯的光谱辐照度标定内部照明的积分球开口光谱辐亮度,这一技术目前已比较成熟,通过两种方式推导出的传递仪器光谱辐亮度响应度之比随波长而变化。针对这一现象,构建了基于标准探测器的绝对辐亮度计和绝对辐照度计。用所构建的绝对辐射计在4个波长处标定光谱辐照度标准灯及标准漫反射板的变化情况,得出标准灯实际光谱辐照度值比自身标定值大,漫反射板实际双向反射分布函数(BRDF)比自身标定BRDF小的结论。将实际光谱辐照度及BRDF分别修正到积分球光谱辐亮度及传递仪器光谱辐亮度响应度中,修正后的传递仪器光谱辐亮度响应度之比随波长变化而趋于常数1。不确定度分析表明,在4个波长处光谱辐亮度定标中合成标准不确定度为0.884%~1.217%,光谱辐照度定标中合成标准不确定度为0.768%~1.136%。
Abstract
In the normal spectral radiance calibration of the internal illumination integrating sphere based on the spectral irradiance of the standard lamp, a deviation changing with wavelength exists in the ratio of instrument′s spectral radiance responsivities derived by two methods. The absolute radiance and irradiance radiometers based on the standard detectors are built up. By calibrating standard lamp of spectral irradiance and standard diffuser with these apparatuses at four wavelengths, it is found that the actual spectral irradiance of the standard lamp is greater while the actual bi-reflection distribution function (BRDF) of the diffuser is smaller than the self-calibrated value. After correcting the standard spectral irradiance and BRDF to the spectral radiance of the integrating sphere and the instrument′s spectral radiance responsivity, the ratio derives to constant 1. Analysis shows a combined standard uncertainty of 0.884%~1.217% for the spectral radiance and 0.768%~1.136% for the spectral irradiance measurement at the four wavelengths.
参考文献

[1] 王锐, 王淑荣, 李福田 等. 真空紫外探测器辐射定标研究[J]. 光学学报, 2010, 30(4): 1026~1030

    Wang Rui, Wang Shurong, Li Futian et al.. Investigation on radiation calibration system of vacuum ultraviolet detector[J]. Acta Optica Sinica, 2010, 30(4): 1026~1030

[2] 李照洲, 郑小兵, 吴浩宇 等. 高精度光谱辐射标准探测器的温度特性研究[J]. 光学学报, 2004, 24(3): 401~407

    Li Zhaozhou, Zheng Xiaobing, Wu Haoyu et al.. Temperature characteristics research of high-precision spectral radiometer standard detector[J]. Acta Optica Sinica, 2004, 24(3): 401~407

[3] 王锐, 宋克非. 高精度紫外探测器辐射定标系统[J]. 光学精密工程, 2009, 17(3): 469~474

    Wang Rui, Song Kefei. High-accuracy radiance calibration system for ultraviolet detector[J]. Optics & Precision Engineering, 2009, 17(3): 469~474

[4] 杨华元, 崔敦杰, 任建伟 等. 基于探测器的成象光谱仪绝对辐射定标方法[J]. 计量学报, 1998, 19(2): 123~129

    Yang Huayuan, Cui Dunjie, Ren Jianwei et al.. Detector based radiometric calibration for imaging spectrometer[J]. Acta Metrologica Sinica, 1998, 19(2): 123~129

[5] Georgi T. Georgiev, James J. Butler. Long-term calibration monitoring of spectral on diffusers BRDF in the air-ultraviolet[J]. Appl. Opt., 2007, 46(32): 7892~7899

[6] S. Janz, E. Hilsenrath, J. Butler et al.. Uncertainties in radiance calibrations of backscatter ultraviolet (BUV) instruments[J]. Metrologia, 1995, 96(32): 637~641

[7] 王锐. 紫外真空紫外探测器定标技术研究[D]. 长春: 中国科学院长春光学精密机械与物理研究所, 2010. 39~55

    Wang Rui. Research on the Calibration Technology of UV-VUV Detector[D]. Changchun: Changchun Institute of Optics, Fine Mechanics and Physics,Chinese Academy of Sciences, 2010. 39~55

[8] James H. Walker, Ambler Thompson. Spectral radiance of a large-area integrating sphere source[J]. J. Res. Natl. Ins. Stand. Technol., 1995, 100: 37~41

[9] Steven W. Brown, Robert D. Saunders, Zhigang Li et al.. An absolute detector-based spectral radiance source[C]. SPIE, 2010, 7807: 78070A

[10] D. F. Heath, Z. Wei, W. K. Fowler et al.. Comparison of spectral radiance calibrations of SBUV-2 satellite ozone monitoring instruments using integrating sphere and flat-plate diffuser techniques[J]. Metrologia, 1993, 30(4): 259~264

[11] J. H. Walker, C. L. Cromer. A technique for improving the calibration of large-area sphere sources[C]. SPIE, 1991, 1493: 224~230

[12] D. F. Heath. Large aperture spectral radiance calibration source for ultraviolet remote sensing instruments[C]. SPIE, 2003, 4891: 335~342

[13] D. F. Heath, Z. Wei. Comparability of spectral radiance calibrations of large aperture earth observing instruments based upon diffuse reflective panels and internally illuminated spherical integrator techniques[C]. SPIE, 1994, 2209: 148~159

[14] D. F. Heath, Z. Wei. Comparisons of UV spectral radiance calibrations of SBUV2 type ozone monitoring instruments and traceability to NIST[C]. SPIE, 1994, 2282: 313~323

[15] 邢进, 王淑荣, 李福田. 空间紫外遥感光谱辐射计光谱辐亮度定标三种方法的比较[J]. 中国激光, 2006, 33(4): 509~515

    Xing Jin, Wang Shurong, Li Futian. Comparison of spectral radiance calibrations of spectroradiometer for ultraviolet space remote sensing using three calibration techniques[J]. Chinese J. Lasers, 2006, 33(4): 509~515

[16] 周磊, 郑小兵. 高精度分光光度计测量光谱透过率[J]. 光电工程, 2006, 33(12): 32~38

    Zhou Lei, Zheng Xiaobing. High-accuracy spectrometer for measurement of spectral transmissivity[J]. Opto-Electronic Engineering, 2006, 33(12): 32~38

杨小虎, 王淑荣, 黄煜, 曲艺, 林冠宇. 基于标准探测器研究标准灯光谱辐照度和漫反射板双向反射分布函数随波长的变化[J]. 光学学报, 2011, 31(6): 0612008. Yang Xiaohu, Wang Shurong, Huang Yu, Qu Yi, Lin Guanyu. Wavelength Dependence of the Spectral Irradiance of Standard Lamp and Bi-Reflection Distribution Function of Diffuser Based on Standard Detectors[J]. Acta Optica Sinica, 2011, 31(6): 0612008.

本文已被 8 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!